Electrostatic nonlinearity limits the use of Capacitive Micromachined Ultrasonic Transducers (CMUTs) in nonlinear ultrasound modalities, such as contrast and tissue harmonic imaging. In this paper, we experimentally investigate the performance of a linearization method based on the appli-cation of pre-distorted analog excitation signals whose spectral content compensates the harmonic distortion introduced by the CMUT. The pre-distortion is based on the accurate knowledge of the CMUT non-linear behavior, achieved by estimating the electrical impedance at several bias voltages. We tested the method using a 256-element, 7.5MHz CMUT linear array probe connected to the ULA-OP 256 imaging system. We measured the acoustic pressure generated in water by a CMUT array element excited with both 4 MHz undistorted and pre-distorted signals. When using the pre-distorted signal, the sec-ond harmonic pressure level was reduced by 13 dB, resulting in a final harmonic-to-fundamental level of -30 dB. Fundamen-tal and second harmonic B-mode images of a commercial tis-sue-mimicking phantom were finally generated using both un-distorted and pre-distorted signals. By computing the second harmonic residual power in a region of interest, a reduction of 13 dB was confirmed using the proposed linearization method in a nonlinear ultrasound imaging context.

Nonlinear ultrasound imaging experiments using a CMUT probe / Savoia, Alessandro Stuart; Caliano, Giosuè; Matrone, Giulia; Ramalli, Alessandro; Boni, Enrico; Tortoli, Piero. - ELETTRONICO. - 2016-:(2016), pp. 1-4. (Intervento presentato al convegno 2016 IEEE International Ultrasonics Symposium, IUS 2016 tenutosi a fra nel 2016) [10.1109/ULTSYM.2016.7728699].

Nonlinear ultrasound imaging experiments using a CMUT probe

RAMALLI, ALESSANDRO;BONI, ENRICO;TORTOLI, PIERO
2016

Abstract

Electrostatic nonlinearity limits the use of Capacitive Micromachined Ultrasonic Transducers (CMUTs) in nonlinear ultrasound modalities, such as contrast and tissue harmonic imaging. In this paper, we experimentally investigate the performance of a linearization method based on the appli-cation of pre-distorted analog excitation signals whose spectral content compensates the harmonic distortion introduced by the CMUT. The pre-distortion is based on the accurate knowledge of the CMUT non-linear behavior, achieved by estimating the electrical impedance at several bias voltages. We tested the method using a 256-element, 7.5MHz CMUT linear array probe connected to the ULA-OP 256 imaging system. We measured the acoustic pressure generated in water by a CMUT array element excited with both 4 MHz undistorted and pre-distorted signals. When using the pre-distorted signal, the sec-ond harmonic pressure level was reduced by 13 dB, resulting in a final harmonic-to-fundamental level of -30 dB. Fundamen-tal and second harmonic B-mode images of a commercial tis-sue-mimicking phantom were finally generated using both un-distorted and pre-distorted signals. By computing the second harmonic residual power in a region of interest, a reduction of 13 dB was confirmed using the proposed linearization method in a nonlinear ultrasound imaging context.
2016
IEEE International Ultrasonics Symposium, IUS
2016 IEEE International Ultrasonics Symposium, IUS 2016
fra
2016
Savoia, Alessandro Stuart; Caliano, Giosuè; Matrone, Giulia; Ramalli, Alessandro; Boni, Enrico; Tortoli, Piero
File in questo prodotto:
File Dimensione Formato  
2016 Savoia - IUS - Nonlinear US imaging experiments using a CMUT probe.pdf

Accesso chiuso

Tipologia: Pdf editoriale (Version of record)
Licenza: Tutti i diritti riservati
Dimensione 3.78 MB
Formato Adobe PDF
3.78 MB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1079112
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 0
social impact