Aluminum oxide based optically stimulated luminescent dosimeters (OSLD) have been recognized as a useful dosimeter for measuring CT dose, particularly for patient dose measurements. Despite the increasing use of this dosimeter, appropriate dosimeter calibration techniques have not been established in the literature; while the manufacturer offers a calibration procedure, it is known to have relatively large uncertainties. The purpose of this work was to evaluate two clinical approaches for calibrating these dosimeters for CT applications, and to determine the uncertainty associated with measurements using these techniques. Three unique calibration procedures were used to calculate dose for a range of CT conditions using a commercially available OSLD and reader. The three calibration procedures included calibration (a) using the vendor‐provided method, (b) relative to a 120 kVp CT spectrum in air, and (c) relative to a megavoltage beam (implemented with 60Co). The dose measured using each of these approaches was compared to dose measured using a calibrated farmer‐type ion chamber. Finally, the uncertainty in the dose measured using each approach was determined. For the CT and megavoltage calibration methods, the dose measured using the OSLD nanoDot was within 5% of the dose measured using an ion chamber for a wide range of different CT scan parameters (80–140 kVp, and with measurements at a range of positions). When calibrated using the vendor‐recommended protocol, the OSLD measured doses were on average 15.5% lower than ion chamber doses. Two clinical calibration techniques have been evaluated and are presented in this work as alternatives to the vendor provided calibration approach. These techniques provide high precision for OSLD based measurements in a CT environment.

Calibration strategies for use of the nanoDot OSLD in CT applications / Scarboro S.B.; Cody D.; Stingo F.C.; Alvarez P.; Followill D.; Court L.; Zhang D.; McNitt-Gray M.; Kry S.F.. - In: JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS. - ISSN 1526-9914. - STAMPA. - 20:(2019), pp. 331-339. [10.1002/acm2.12491]

Calibration strategies for use of the nanoDot OSLD in CT applications

Stingo F. C.;SANCHEZ ALVAREZ, CLAUDIA PAZ;
2019

Abstract

Aluminum oxide based optically stimulated luminescent dosimeters (OSLD) have been recognized as a useful dosimeter for measuring CT dose, particularly for patient dose measurements. Despite the increasing use of this dosimeter, appropriate dosimeter calibration techniques have not been established in the literature; while the manufacturer offers a calibration procedure, it is known to have relatively large uncertainties. The purpose of this work was to evaluate two clinical approaches for calibrating these dosimeters for CT applications, and to determine the uncertainty associated with measurements using these techniques. Three unique calibration procedures were used to calculate dose for a range of CT conditions using a commercially available OSLD and reader. The three calibration procedures included calibration (a) using the vendor‐provided method, (b) relative to a 120 kVp CT spectrum in air, and (c) relative to a megavoltage beam (implemented with 60Co). The dose measured using each of these approaches was compared to dose measured using a calibrated farmer‐type ion chamber. Finally, the uncertainty in the dose measured using each approach was determined. For the CT and megavoltage calibration methods, the dose measured using the OSLD nanoDot was within 5% of the dose measured using an ion chamber for a wide range of different CT scan parameters (80–140 kVp, and with measurements at a range of positions). When calibrated using the vendor‐recommended protocol, the OSLD measured doses were on average 15.5% lower than ion chamber doses. Two clinical calibration techniques have been evaluated and are presented in this work as alternatives to the vendor provided calibration approach. These techniques provide high precision for OSLD based measurements in a CT environment.
2019
20
331
339
Scarboro S.B.; Cody D.; Stingo F.C.; Alvarez P.; Followill D.; Court L.; Zhang D.; McNitt-Gray M.; Kry S.F.
File in questo prodotto:
File Dimensione Formato  
Scarboro_JAMP2018.pdf

Accesso chiuso

Tipologia: Pdf editoriale (Version of record)
Licenza: Tutti i diritti riservati
Dimensione 402.34 kB
Formato Adobe PDF
402.34 kB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1171112
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 7
social impact