BACKGROUND: Two point-of-care (POC) systems have been recently proposed as rapid tools with which to evaluate residual platelet reactivity (RPR) in coronary artery disease (CAD) patients. OBJECTIVES AND METHODS: We compared Platelet Function Analyzer-100 (PFA-100) closure times (CTs) by collagen/adenosine 5'-diphosphate (ADP) (C/ADP CT) cartridge and the VerifyNow P2Y12 Assay (VerifyNow) with light transmission aggregation (LTA) induced by 2 and 10 micromol L(-1) ADP in 1267 CAD patients on dual antiplatelet therapy who underwent percutaneous coronary intervention. We also performed the vasodilator-stimulated phosphoprotein (VASP) phosphorylation assay by cytofluorimetric analysis in a subgroup of 115 patients. RESULTS: Cut-off values for identifying RPR were: > or = 54% and > or = 66% for LTA induced by 2 and 10 micromol L(-1) ADP respectively, and > or = 264 P2Y12 Reaction Units (PRU) for VerifyNow. The cut-off for PFA-100 C/ADP CT was > or = 68 s. RPR was detected in 25.1% of patients by 2 mumol L(-1) ADP-induced LTA (ADP-LTA), in 23.2% by 10 micromol L(-1) ADP-LTA, in 24.4% by PFA-100, and in 24.7% by VerifyNow. PFA-100 results did not parallel those obtained with LTA. VerifyNow showed a significant correlation (rho = 0.62, P < 0.001) and significant agreement (k = 0.34, P < 0.001) with LTA induced by 2 micromol L(-1) ADP. The correlation was similar but the agreement was better between VerifyNow and 10 micromol L(-1) ADP-LTA (rho = 0.64, P < 0.0001; k = 0.43, P < 0.001). Significant relationships were found between VASP platelet reactivity index and both ADP-LTA and VerifyNow. PFA-100 C/ADP CT did not significantly correlate with any of the other assays. CONCLUSIONS: Our results show a significant correlation between LTA and VerifyNow but not the PFA-100 C/ADP assay. Clinical validation studies for POC systems are necessary.

Different methodologies for evaluating the effect of clopidogrel on platelet function in high-risk coronary artery disease patients / R.Paniccia; E.Antonucci; A.M.Gori; R.Marcucci; C.Giglioli; D.Antoniucci; G.F.Gensini; R.Abbate; D.Prisco. - In: JOURNAL OF THROMBOSIS AND HAEMOSTASIS. - ISSN 1538-7933. - STAMPA. - 5:(2007), pp. 1839-1847.

Different methodologies for evaluating the effect of clopidogrel on platelet function in high-risk coronary artery disease patients

PANICCIA, RITA;GORI, ANNA MARIA;MARCUCCI, ROSSELLA;GENSINI, GIAN FRANCO;ABBATE, ROSANNA;PRISCO, DOMENICO
2007

Abstract

BACKGROUND: Two point-of-care (POC) systems have been recently proposed as rapid tools with which to evaluate residual platelet reactivity (RPR) in coronary artery disease (CAD) patients. OBJECTIVES AND METHODS: We compared Platelet Function Analyzer-100 (PFA-100) closure times (CTs) by collagen/adenosine 5'-diphosphate (ADP) (C/ADP CT) cartridge and the VerifyNow P2Y12 Assay (VerifyNow) with light transmission aggregation (LTA) induced by 2 and 10 micromol L(-1) ADP in 1267 CAD patients on dual antiplatelet therapy who underwent percutaneous coronary intervention. We also performed the vasodilator-stimulated phosphoprotein (VASP) phosphorylation assay by cytofluorimetric analysis in a subgroup of 115 patients. RESULTS: Cut-off values for identifying RPR were: > or = 54% and > or = 66% for LTA induced by 2 and 10 micromol L(-1) ADP respectively, and > or = 264 P2Y12 Reaction Units (PRU) for VerifyNow. The cut-off for PFA-100 C/ADP CT was > or = 68 s. RPR was detected in 25.1% of patients by 2 mumol L(-1) ADP-induced LTA (ADP-LTA), in 23.2% by 10 micromol L(-1) ADP-LTA, in 24.4% by PFA-100, and in 24.7% by VerifyNow. PFA-100 results did not parallel those obtained with LTA. VerifyNow showed a significant correlation (rho = 0.62, P < 0.001) and significant agreement (k = 0.34, P < 0.001) with LTA induced by 2 micromol L(-1) ADP. The correlation was similar but the agreement was better between VerifyNow and 10 micromol L(-1) ADP-LTA (rho = 0.64, P < 0.0001; k = 0.43, P < 0.001). Significant relationships were found between VASP platelet reactivity index and both ADP-LTA and VerifyNow. PFA-100 C/ADP CT did not significantly correlate with any of the other assays. CONCLUSIONS: Our results show a significant correlation between LTA and VerifyNow but not the PFA-100 C/ADP assay. Clinical validation studies for POC systems are necessary.
2007
5
1839
1847
R.Paniccia; E.Antonucci; A.M.Gori; R.Marcucci; C.Giglioli; D.Antoniucci; G.F.Gensini; R.Abbate; D.Prisco
File in questo prodotto:
File Dimensione Formato  
different methodologies.pdf

Accesso chiuso

Tipologia: Versione finale referata (Postprint, Accepted manuscript)
Licenza: Tutti i diritti riservati
Dimensione 730.92 kB
Formato Adobe PDF
730.92 kB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/318827
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 173
  • ???jsp.display-item.citation.isi??? 153
social impact