In freehand elastography, quasi-static tissue compression is applied through the ultrasound probe, and the corresponding axial strain is estimated by calculating the time shift between consecutive echo signals. This calculation typically suffers from a poor signal-to-noise ratio or from the decorrelation between consecutive echoes resulting from an erroneous axial motion impressed by the operator. This paper shows that the quality of elastograms can be improved through the integration of two distinct techniques in the strain estimation procedure. The first technique evaluates the displacement of the tissue by analyzing the phases of the echo signal spectra acquired during compression. The second technique increases the displacement estimation robustness by averaging multiple displacement estimations in a high-frame-rate imaging system, while maintaining the typical elastogram frame-rate. The experimental results, obtained with the Ultrasound Advanced Open Platform (ULA-OP) and a cyst phantom, demonstrate that each of the proposed methods can independently improve the quality of elastograms, and that further improvements are possible through their combination.

Frequency domain-based strain estimation and high frame-rate imaging for quasi-static elastography / A.Ramalli; O.Basset; C.Cachard; E.Boni; P.Tortoli. - In: IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL. - ISSN 0885-3010. - STAMPA. - 59:(2012), pp. 817-824. [10.1109/TUFFC.2012.2260]

Frequency domain-based strain estimation and high frame-rate imaging for quasi-static elastography

RAMALLI, ALESSANDRO;BONI, ENRICO;TORTOLI, PIERO
2012

Abstract

In freehand elastography, quasi-static tissue compression is applied through the ultrasound probe, and the corresponding axial strain is estimated by calculating the time shift between consecutive echo signals. This calculation typically suffers from a poor signal-to-noise ratio or from the decorrelation between consecutive echoes resulting from an erroneous axial motion impressed by the operator. This paper shows that the quality of elastograms can be improved through the integration of two distinct techniques in the strain estimation procedure. The first technique evaluates the displacement of the tissue by analyzing the phases of the echo signal spectra acquired during compression. The second technique increases the displacement estimation robustness by averaging multiple displacement estimations in a high-frame-rate imaging system, while maintaining the typical elastogram frame-rate. The experimental results, obtained with the Ultrasound Advanced Open Platform (ULA-OP) and a cyst phantom, demonstrate that each of the proposed methods can independently improve the quality of elastograms, and that further improvements are possible through their combination.
2012
59
817
824
A.Ramalli; O.Basset; C.Cachard; E.Boni; P.Tortoli
File in questo prodotto:
File Dimensione Formato  
Elastography_Abs.pdf

accesso aperto

Tipologia: Altro
Licenza: Tutti i diritti riservati
Dimensione 23.37 kB
Formato Adobe PDF
23.37 kB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/596102
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 26
  • ???jsp.display-item.citation.isi??? 24
social impact