Tubular potassium-selective flow-through electrodes were obtained by treating silicone-rubber tubing with a solution of valinomycin in chloroform. They were compared with conventional PVC-based electrodes in a flow-through cell. When serum and whole blood were pumped continuously for 30 min and 5 h, the response and wash times for the electrodes were found to vary considerably. The silicone rubber-based potassium sensor showed better biocompatibility with virtually unchanged dynamic features after prolonged contact with blood.

Improved potentiometric determination of potassium in whole blood and serum with a valinomycin-treated silicone-rubber tubular electrode / Marco, Mascini; Giovanna, Marrazza. - In: ANALYTICA CHIMICA ACTA. - ISSN 0003-2670. - STAMPA. - 231(1):(1990), pp. 125-128.

Improved potentiometric determination of potassium in whole blood and serum with a valinomycin-treated silicone-rubber tubular electrode

MASCINI, MARCO;MARRAZZA, GIOVANNA
1990

Abstract

Tubular potassium-selective flow-through electrodes were obtained by treating silicone-rubber tubing with a solution of valinomycin in chloroform. They were compared with conventional PVC-based electrodes in a flow-through cell. When serum and whole blood were pumped continuously for 30 min and 5 h, the response and wash times for the electrodes were found to vary considerably. The silicone rubber-based potassium sensor showed better biocompatibility with virtually unchanged dynamic features after prolonged contact with blood.
1990
231(1)
125
128
Marco, Mascini; Giovanna, Marrazza
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/650865
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 19
social impact