The aim of this work is to assess the accuracy of a surface matching registration (SMR) technique for the correlation of cardiac studies in positron emission tomography (PET). Registration parameters were estimated by matching corresponding body surfaces, extracted from transmission studies, aligned to the PET emission images to be correlated. The accuracy of the SMR technique in this specific application was assessed by computer simulations, phantom experiments and on clinical PET data. Registration accuracy was evaluated in relation to the body surfaces (external, internal and the combination of the two) used by the SMR method. Better results were found when matching shaped and irregular surfaces such as internal lung contours. The robustness of the method was verified for different counting statistics recorded in transmission images. A clinical validation of the SMR method was performed on fluorine-18-deoxyglucose PET cardiac studies.

APPLICATION OF A SURFACE MATCHING IMAGE REGISTRATION TECHNIQUE TO THE CORRELATION OF CARDIAC STUDIES IN POSITRON EMISSION TOMOGRAPHY (PET) BY TRANSMISSION IMAGES / S. PALLOTTA; M. GILARDI; V. BETTINARDI; G. RIZZO; C. LANDONI; G. STRIANO; R. MASI; F. FAZIO. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - ELETTRONICO. - 40:(1995), pp. 1695-1708. [10.1088/0031-9155/40/10/010]

APPLICATION OF A SURFACE MATCHING IMAGE REGISTRATION TECHNIQUE TO THE CORRELATION OF CARDIAC STUDIES IN POSITRON EMISSION TOMOGRAPHY (PET) BY TRANSMISSION IMAGES

PALLOTTA, STEFANIA;
1995

Abstract

The aim of this work is to assess the accuracy of a surface matching registration (SMR) technique for the correlation of cardiac studies in positron emission tomography (PET). Registration parameters were estimated by matching corresponding body surfaces, extracted from transmission studies, aligned to the PET emission images to be correlated. The accuracy of the SMR technique in this specific application was assessed by computer simulations, phantom experiments and on clinical PET data. Registration accuracy was evaluated in relation to the body surfaces (external, internal and the combination of the two) used by the SMR method. Better results were found when matching shaped and irregular surfaces such as internal lung contours. The robustness of the method was verified for different counting statistics recorded in transmission images. A clinical validation of the SMR method was performed on fluorine-18-deoxyglucose PET cardiac studies.
1995
40
1695
1708
S. PALLOTTA; M. GILARDI; V. BETTINARDI; G. RIZZO; C. LANDONI; G. STRIANO; R. MASI; F. FAZIO
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/779784
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 20
  • ???jsp.display-item.citation.isi??? 13
social impact