A prototype of bidimensional dosimeter made on a 2.5×2.5 cm2 active area polycrystalline Chemical Vapour Deposited (pCVD) diamond film, equipped with a matrix of 12×12 contacts connected to the read-out electronics, has been used to evaluate a map of dose under Intensity Modulated Radiation Therapy (IMRT) fields for a possible application in pre-treatment verifications of cancer treatments. Tests have been performed under a 6-10 MVRX beams with IMRT fields for prostate and breast cancer. Measurements have been taken by measuring the 144 pixels in different positions, obtained by shifting the device along the x/y axes to span a total map of 14.4×10 cm2. Results show that absorbed doses measured by our pCVD diamond device are consistent with those calculated by the Treatment Planning System (TPS).
First dose-map measured with a polycrystalline diamond 2D dosimeter under an intensity modulated radiotherapy beam / Scaringella, M; Zani, M.; Baldi, A.; Bucciolini, M.; Pace, E.; De Sio, A.; Talamonti, C.; Bruzzi, M.. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - STAMPA. - 796:(2015), pp. 89-92. [10.1016/j.nima.2015.03.023]
First dose-map measured with a polycrystalline diamond 2D dosimeter under an intensity modulated radiotherapy beam
SCARINGELLA, MONICA;BUCCIOLINI, MARTA;PACE, EMANUELE;TALAMONTI, CINZIA;BRUZZI, MARA
2015
Abstract
A prototype of bidimensional dosimeter made on a 2.5×2.5 cm2 active area polycrystalline Chemical Vapour Deposited (pCVD) diamond film, equipped with a matrix of 12×12 contacts connected to the read-out electronics, has been used to evaluate a map of dose under Intensity Modulated Radiation Therapy (IMRT) fields for a possible application in pre-treatment verifications of cancer treatments. Tests have been performed under a 6-10 MVRX beams with IMRT fields for prostate and breast cancer. Measurements have been taken by measuring the 144 pixels in different positions, obtained by shifting the device along the x/y axes to span a total map of 14.4×10 cm2. Results show that absorbed doses measured by our pCVD diamond device are consistent with those calculated by the Treatment Planning System (TPS).I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.