This study aims at investigating the possible employment for pre-treatment verifications in Intensity Modulated Radiation Therapy (IMRT) of a polycrystalline Chemical Vapour Deposited (pCVD) diamond bidimensional detector. The pCVD device, with an area of 2.5 ×2.5cm2 (12× 12 pixels), has been used to measure dose maps of a 10 MVRX prostatic IMRT field. Its response was compared both with a commercial bi-dimensional detector made with silicon and with Treatment Planning System (TPS) calculations. Measurement provided promising results on a map of 1.8×12.6cm2. Absorbed doses measured along IMRT profiles by our device are consistent with the ones acquired with the commercial device and an overall good agreement with respect to the TPS was found for the diamond dosimeter.
Bidimensional polycrystalline CVD diamond detector for Intensity Modulated Radiation Therapy pre-treatment verifications / Zani, M.; Scaringella, M.; Talamonti, C.; De Sio, A.; Pace, E.; Tozzetti, L.; Baldi, A.; Bucciolini, M.; Bruzzi, M.. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - ELETTRONICO. - 10:(2015), pp. 0-0. [10.1088/1748-0221/10/03/C03046]
Bidimensional polycrystalline CVD diamond detector for Intensity Modulated Radiation Therapy pre-treatment verifications
ZANI, MARGHERITA;SCARINGELLA, MONICA;TALAMONTI, CINZIA;DE SIO, ANTONIO;PACE, EMANUELE;BALDI, ANDREA;BUCCIOLINI, MARTA;BRUZZI, MARA
2015
Abstract
This study aims at investigating the possible employment for pre-treatment verifications in Intensity Modulated Radiation Therapy (IMRT) of a polycrystalline Chemical Vapour Deposited (pCVD) diamond bidimensional detector. The pCVD device, with an area of 2.5 ×2.5cm2 (12× 12 pixels), has been used to measure dose maps of a 10 MVRX prostatic IMRT field. Its response was compared both with a commercial bi-dimensional detector made with silicon and with Treatment Planning System (TPS) calculations. Measurement provided promising results on a map of 1.8×12.6cm2. Absorbed doses measured along IMRT profiles by our device are consistent with the ones acquired with the commercial device and an overall good agreement with respect to the TPS was found for the diamond dosimeter.I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.