The objective of this work was to assess both the perception of failure modes in Intensity Modulated Radiation Therapy (IMRT) when the linac is operated at the edge of tolerances given in AAPM TG-40 (Kutcher et al.) and TG-142 (Klein et al.) as well as the application of FMEA to this specific section of the IMRT process
An FMEA evaluation of intensity modulated radiation therapy dose delivery failures at tolerance criteria levels / Faught, Jacqueline Tonigan; Balter, Peter A.; Johnson, Jennifer L.; Kry, Stephen F.; Court, Laurence E.; Stingo, FRANCESCO CLAUDIO; Followill, David S.. - In: MEDICAL PHYSICS. - ISSN 0094-2405. - STAMPA. - 44:(2017), pp. 5575-5583. [10.1002/mp.12551]
An FMEA evaluation of intensity modulated radiation therapy dose delivery failures at tolerance criteria levels
Stingo, Francesco C.;
2017
Abstract
The objective of this work was to assess both the perception of failure modes in Intensity Modulated Radiation Therapy (IMRT) when the linac is operated at the edge of tolerances given in AAPM TG-40 (Kutcher et al.) and TG-142 (Klein et al.) as well as the application of FMEA to this specific section of the IMRT processFile | Dimensione | Formato | |
---|---|---|---|
Faught_et_al-2017-Medical_Physics.pdf
Accesso chiuso
Tipologia:
Pdf editoriale (Version of record)
Licenza:
DRM non definito
Dimensione
499.91 kB
Formato
Adobe PDF
|
499.91 kB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.