PURPOSE: The 11-gauge (11G) stereotactic vacuum-assisted breast biopsy (VABB) showed a better profile than 14G-VABB in terms of feasibility, safety, microcalcification sampling, and accuracy. Underestimation rates were significantly lower with 11G-VABB than with 14G-VABB. Thus, the introduction of an even larger needle at the VABB procedure could reduce this rate further. The purpose of this study was to compare the overall performance of stereotactic VABB with 8G and 11G needles. MATERIALS AND METHODS: Four hundred and three VABBs performed between July 2012 and February 2015 at the Breast Diagnostic Unit of Careggi Hospital in Florence were retrospectively analyzed; 197 were performed with 11G-VABB and 206 with 8G-VABB. Lesions were classified according to mammographical patterns in microcalcifications, architectural distortions, or opacities, and all biopsy targets were classified according to BIRADS classification as BIRADS III, IV or V. Data were collected on radiological classification of targets, imaging presentation, procedure time, number of specimens per procedure, and microcalcification retrieval on histological findings. Surgery was always performed when high-risk or malignant lesions (B3 or B5) were detected; the final diagnosis was made on surgical pathology. RESULTS: Compared to VABB with an 11G needle, 8G-VABB allows a reduction in the time needed to complete the procedure (20.6 versus 27.4, P < 0.00001) and the number of specimens collected per lesion (21.6 versus 12.2, P < 0.00001). Moreover, 8G-VABB resulted in the same diagnostic accuracy, and the underestimation rates were comparable between the two groups for both B3 and DCIS lesions. CONCLUSIONS: The 8G needle should be considered as a valid alternative option in VABB for breast lesions.
Stereotactic vacuum-assisted breast biopsy: Comparison between 11- and 8-gauge needles / Ruggirello, Irene; Nori, Jacopo; Desideri, Isacco; Saieva, Calogero; Giannotti, Elisabetta; Bicchierai, Giulia; De Benedetto, Diego; Francolini, Giulio; Bianchi, Simonetta; Vezzosi, Vania; Sanchez, Luis; Susini, Tommaso; Orzalesi, Lorenzo; Meattini, Icro; Livi, Lorenzo; Miele, Vittorio. - In: EUROPEAN JOURNAL OF SURGICAL ONCOLOGY. - ISSN 0748-7983. - STAMPA. - 43:(2017), pp. 2257-2260. [10.1016/j.ejso.2017.07.011]
Stereotactic vacuum-assisted breast biopsy: Comparison between 11- and 8-gauge needles
Desideri, Isacco
Writing – Review & Editing
;Bicchierai, Giulia;De Benedetto, Diego;Francolini, Giulio;Bianchi, Simonetta;Vezzosi, Vania;Susini, Tommaso;Orzalesi, Lorenzo;Meattini, Icro;Livi, Lorenzo;Miele, Vittorio
2017
Abstract
PURPOSE: The 11-gauge (11G) stereotactic vacuum-assisted breast biopsy (VABB) showed a better profile than 14G-VABB in terms of feasibility, safety, microcalcification sampling, and accuracy. Underestimation rates were significantly lower with 11G-VABB than with 14G-VABB. Thus, the introduction of an even larger needle at the VABB procedure could reduce this rate further. The purpose of this study was to compare the overall performance of stereotactic VABB with 8G and 11G needles. MATERIALS AND METHODS: Four hundred and three VABBs performed between July 2012 and February 2015 at the Breast Diagnostic Unit of Careggi Hospital in Florence were retrospectively analyzed; 197 were performed with 11G-VABB and 206 with 8G-VABB. Lesions were classified according to mammographical patterns in microcalcifications, architectural distortions, or opacities, and all biopsy targets were classified according to BIRADS classification as BIRADS III, IV or V. Data were collected on radiological classification of targets, imaging presentation, procedure time, number of specimens per procedure, and microcalcification retrieval on histological findings. Surgery was always performed when high-risk or malignant lesions (B3 or B5) were detected; the final diagnosis was made on surgical pathology. RESULTS: Compared to VABB with an 11G needle, 8G-VABB allows a reduction in the time needed to complete the procedure (20.6 versus 27.4, P < 0.00001) and the number of specimens collected per lesion (21.6 versus 12.2, P < 0.00001). Moreover, 8G-VABB resulted in the same diagnostic accuracy, and the underestimation rates were comparable between the two groups for both B3 and DCIS lesions. CONCLUSIONS: The 8G needle should be considered as a valid alternative option in VABB for breast lesions.File | Dimensione | Formato | |
---|---|---|---|
2017_Ruggirello et al_EJSO_breast biopsy 11 vs 18.pdf
Accesso chiuso
Descrizione: Paper
Tipologia:
Pdf editoriale (Version of record)
Licenza:
Tutti i diritti riservati
Dimensione
152.96 kB
Formato
Adobe PDF
|
152.96 kB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.