BACKGROUND: Aortic prosthesis area (EOA) is computed by continuity equation from left ventricular (LV) stroke volume (SV) derived from LV outflow tract diameter (LVOTD ) or, when unmeasurable, from LV volumes (SVV ). There is evidence to suggest LVOT ellipticity and recommend 3D LVOT area (LVOTCSA ) adoption in aortic stenosis. We sought to evaluate if the same concept applies to supra-annular aortic prosthesis comparing SV and EOA derived from LVOTD (EOAD ) and from LVOTCSA (EOACSA ). EOA computed from SVV (EAOV ) accuracy was evaluated in this setting. Patient-prosthesis mismatch (PPM) was compared among different EOA computations. METHODS: A consecutive series of 202 patients (aged 81 ± 4 years, 43% males) underwent St.Jude Trifecta aortic valve replacement (AVR) and were followed up with echocardiography at one-year (335 ± 31 days). All measurements followed the EACVI or ASE guidelines, 3D X-plane modality was used to compute SVv and measure LVOTCSA ; SV was calculated from LVOTD (SVD ) and LVOTCSA (SVCSA ). PPM was indexed EOA <0.65 cm²/m². RESULTS: LVOT showed a significant ellipticity index (1.17 ± .27), independent of prosthesis size. EOAD (1.70 ± 0.55 cm²) was less than EOACSA (1.95 ± 0.62 cm²) (P < .0001). SVV was significantly lower than SVD and SVCSA . Bland-Altman analysis showed a significant correlation between SVV and SVD or SVCSA although with large bias and imprecision. The correlations improved reducing bias and imprecision when LVOT time-velocity integral was <20 cm. PPM incidence was higher in EOAV (15.6%) compared to EOAD (P = .04) or EOACSA (P < .001). CONCLUSIONS: In supra-annular AVR, LVOT retains its elliptical shape and LVOTCSA yielded larger prosthesis EOA with lower PPM incidence. PPM may be overestimated by EOAV .

Left ventricular outflow tract shape after aortic valve replacement with St. Jude Trifecta prosthesis / Barletta, Giuseppe*; Venditti, Francesco; Stefano, Pierluigi; Del Bene, Riccarda; Di Mario, Carlo. - In: ECHOCARDIOGRAPHY. - ISSN 0742-2822. - ELETTRONICO. - Dec 22:(2017), pp. 0-0. [10.1111/echo.13778]

Left ventricular outflow tract shape after aortic valve replacement with St. Jude Trifecta prosthesis

Barletta, Giuseppe;VENDITTI, FRANCESCO;Stefano, Pierluigi;Di Mario, Carlo
2017

Abstract

BACKGROUND: Aortic prosthesis area (EOA) is computed by continuity equation from left ventricular (LV) stroke volume (SV) derived from LV outflow tract diameter (LVOTD ) or, when unmeasurable, from LV volumes (SVV ). There is evidence to suggest LVOT ellipticity and recommend 3D LVOT area (LVOTCSA ) adoption in aortic stenosis. We sought to evaluate if the same concept applies to supra-annular aortic prosthesis comparing SV and EOA derived from LVOTD (EOAD ) and from LVOTCSA (EOACSA ). EOA computed from SVV (EAOV ) accuracy was evaluated in this setting. Patient-prosthesis mismatch (PPM) was compared among different EOA computations. METHODS: A consecutive series of 202 patients (aged 81 ± 4 years, 43% males) underwent St.Jude Trifecta aortic valve replacement (AVR) and were followed up with echocardiography at one-year (335 ± 31 days). All measurements followed the EACVI or ASE guidelines, 3D X-plane modality was used to compute SVv and measure LVOTCSA ; SV was calculated from LVOTD (SVD ) and LVOTCSA (SVCSA ). PPM was indexed EOA <0.65 cm²/m². RESULTS: LVOT showed a significant ellipticity index (1.17 ± .27), independent of prosthesis size. EOAD (1.70 ± 0.55 cm²) was less than EOACSA (1.95 ± 0.62 cm²) (P < .0001). SVV was significantly lower than SVD and SVCSA . Bland-Altman analysis showed a significant correlation between SVV and SVD or SVCSA although with large bias and imprecision. The correlations improved reducing bias and imprecision when LVOT time-velocity integral was <20 cm. PPM incidence was higher in EOAV (15.6%) compared to EOAD (P = .04) or EOACSA (P < .001). CONCLUSIONS: In supra-annular AVR, LVOT retains its elliptical shape and LVOTCSA yielded larger prosthesis EOA with lower PPM incidence. PPM may be overestimated by EOAV .
2017
Dec 22
0
0
Goal 3: Good health and well-being for people
Barletta, Giuseppe*; Venditti, Francesco; Stefano, Pierluigi; Del Bene, Riccarda; Di Mario, Carlo
File in questo prodotto:
File Dimensione Formato  
Barletta G- Echocardiography 2017.pdf

Accesso chiuso

Tipologia: Pdf editoriale (Version of record)
Licenza: Tutti i diritti riservati
Dimensione 814.64 kB
Formato Adobe PDF
814.64 kB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1113959
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 7
social impact