BACKGROUND AND PURPOSE: Reported plan quality improvements with autoplanning of radiotherapy of the prostate and seminal vesicles are poor. A system for automated multi-criterial planning has been validated for this treatment in a large international multi-center study. The system is configured with training plans using a mechanism that strives for quality improvements relative to those plans. MATERIAL AND METHODS: Each of the four participating centers included thirty manually generated clinical Volumetric Modulated Arc Therapy prostate plans (manVMAT). Ten plans were used for autoplanning training. The other twenty were compared with an automatically generated plan (autoVMAT). Plan evaluations considered dosimetric plan parameters and blinded side-by-side plan comparisons by clinicians. RESULTS: With equivalent Planning Target Volume (PTV) V95%, D2%, D98%, and dose homogeneity autoVMAT was overall superior for rectum with median differences of 3.4 Gy (p < 0.001) in Dmean, 4.0% (p < 0.001) in V60Gy, and 1.5% (p = 0.001) in V75Gy, and for bladder Dmean (0.9 Gy, p < 0.001). Also the clinicians' plan comparisons pointed at an overall preference for autoVMAT. Advantages of autoVMAT were highly treatment center- and patient-specific with overall ranges for differences in rectum Dmean and V60Gy of [-4,12] Gy and [-2,15]%, respectively. CONCLUSION: Observed advantages of autoplanning were clinically relevant and larger than reported in the literature. The latter is likely related to the multi-criterial nature of the applied autoplanning algorithm, with for each center a dedicated configuration that aims at plan improvements relative to its (clinical) training plans. Large variations among patients in differences between manVMAT and autoVMAT point at inconsistencies in manual planning.
Fully automated, multi-criterial planning for Volumetric Modulated Arc Therapy – An international multi-center validation for prostate cancer / Heijmen B, Voet P, Fransen D, Penninkhof J, Milder M, Akhiat H, Bonomo Pierluigi, Casati Marta, Georg D, Goldner G, Henry A, Lilley J, Lohr F, Marrazzo Livia, Pallotta Stefania, Pellegrini R, Seppenwoolde Y, Simontacchi Gabriele, Steil V, Stieler F, Wilson S, Breedveld S,. - In: RADIOTHERAPY AND ONCOLOGY. - ISSN 0167-8140. - ELETTRONICO. - (2018), pp. 343-348.
Fully automated, multi-criterial planning for Volumetric Modulated Arc Therapy – An international multi-center validation for prostate cancer
Bonomo PierluigiValidation
;Casati MartaValidation
;Marrazzo LiviaFormal Analysis
;Pallotta StefaniaFormal Analysis
;Simontacchi GabrieleValidation
;
2018
Abstract
BACKGROUND AND PURPOSE: Reported plan quality improvements with autoplanning of radiotherapy of the prostate and seminal vesicles are poor. A system for automated multi-criterial planning has been validated for this treatment in a large international multi-center study. The system is configured with training plans using a mechanism that strives for quality improvements relative to those plans. MATERIAL AND METHODS: Each of the four participating centers included thirty manually generated clinical Volumetric Modulated Arc Therapy prostate plans (manVMAT). Ten plans were used for autoplanning training. The other twenty were compared with an automatically generated plan (autoVMAT). Plan evaluations considered dosimetric plan parameters and blinded side-by-side plan comparisons by clinicians. RESULTS: With equivalent Planning Target Volume (PTV) V95%, D2%, D98%, and dose homogeneity autoVMAT was overall superior for rectum with median differences of 3.4 Gy (p < 0.001) in Dmean, 4.0% (p < 0.001) in V60Gy, and 1.5% (p = 0.001) in V75Gy, and for bladder Dmean (0.9 Gy, p < 0.001). Also the clinicians' plan comparisons pointed at an overall preference for autoVMAT. Advantages of autoVMAT were highly treatment center- and patient-specific with overall ranges for differences in rectum Dmean and V60Gy of [-4,12] Gy and [-2,15]%, respectively. CONCLUSION: Observed advantages of autoplanning were clinically relevant and larger than reported in the literature. The latter is likely related to the multi-criterial nature of the applied autoplanning algorithm, with for each center a dedicated configuration that aims at plan improvements relative to its (clinical) training plans. Large variations among patients in differences between manVMAT and autoVMAT point at inconsistencies in manual planning.File | Dimensione | Formato | |
---|---|---|---|
Fully automated, multi criterial planning for Volumetric Modulated Arc.pdf
Accesso chiuso
Tipologia:
Pdf editoriale (Version of record)
Licenza:
Tutti i diritti riservati
Dimensione
663.8 kB
Formato
Adobe PDF
|
663.8 kB | Adobe PDF | Richiedi una copia |
2158-1153036_postprint.pdf
Open Access dal 01/01/2020
Tipologia:
Versione finale referata (Postprint, Accepted manuscript)
Licenza:
Creative commons
Dimensione
151.25 kB
Formato
Adobe PDF
|
151.25 kB | Adobe PDF |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.