In time-domain diffuse optics the sensitivity to localized absorption changes buried inside a diffusive medium depends strongly on the interplay between instrumental, optical and geometrical parameters, which can hinder the theoretical advantages of novel measurement strategies like the short source-detector distance approach. Here, we present a study based on experimental measurements and simulations to comprehensively evaluate the effect of all different parameters. Results are evaluated exploiting standardized figures of merit, like contrast and contrast-to-noise ratio, to quantify the system sensitivity to deep localized absorption perturbations. Key findings show that the most critical hardware parameter is the memory effect which ultimately limits the dynamic range. Further, a choice of the source-detector distance around 10 mm seems to be a good compromise to compensate non-idealities in practical systems still preserving the advantages of short distances. This work provides both indications for users about the best measurement conditions and strategies, and for technology developers to identify the most crucial hardware features in view of next generation diffuse optics systems.

Instrumental, optical and geometrical parameters affecting time-gated diffuse optical measurements: A systematic study / Behera A.; Di Sieno L.; Pifferi A.; Martelli F.; Mora A.D.. - In: BIOMEDICAL OPTICS EXPRESS. - ISSN 2156-7085. - ELETTRONICO. - 9:(2018), pp. 5524-5542. [10.1364/BOE.9.005524]

Instrumental, optical and geometrical parameters affecting time-gated diffuse optical measurements: A systematic study

Martelli F.
Membro del Collaboration Group
;
2018

Abstract

In time-domain diffuse optics the sensitivity to localized absorption changes buried inside a diffusive medium depends strongly on the interplay between instrumental, optical and geometrical parameters, which can hinder the theoretical advantages of novel measurement strategies like the short source-detector distance approach. Here, we present a study based on experimental measurements and simulations to comprehensively evaluate the effect of all different parameters. Results are evaluated exploiting standardized figures of merit, like contrast and contrast-to-noise ratio, to quantify the system sensitivity to deep localized absorption perturbations. Key findings show that the most critical hardware parameter is the memory effect which ultimately limits the dynamic range. Further, a choice of the source-detector distance around 10 mm seems to be a good compromise to compensate non-idealities in practical systems still preserving the advantages of short distances. This work provides both indications for users about the best measurement conditions and strategies, and for technology developers to identify the most crucial hardware features in view of next generation diffuse optics systems.
2018
9
5524
5542
Goal 3: Good health and well-being for people
Behera A.; Di Sieno L.; Pifferi A.; Martelli F.; Mora A.D.
File in questo prodotto:
File Dimensione Formato  
Behera_et_al_boe-9-11-5524.pdf

Accesso chiuso

Descrizione: Articolo principale
Tipologia: Pdf editoriale (Version of record)
Licenza: Tutti i diritti riservati
Dimensione 4.6 MB
Formato Adobe PDF
4.6 MB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1193318
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 23
  • ???jsp.display-item.citation.isi??? 20
social impact