We evaluate the combined usage of two systems, the Integral Quality Monitor (IQM) transmission detector and SoftDiso software, for in vivo dose monitoring by simultaneous detection of delivery and patient setup errors in whole breast irradiation. An Alderson RANDO phantom was adapted with silicon breast prostheses to mimic the female anatomy. Plans with simulated delivery errors were created from a reference left breast plan, and patient setup errors were simulated by moving the phantom. Deviations from reference values recorded by both monitoring systems were measured for all plans and phantom positions. A 2D global gamma analysis was performed in SoftDiso for all phantom displacements. Both IQM signals and SoftDiso R-values are sensitive to small MU variations. However, only IQM is sensitive to jaw position variations. Conversely, IQM is unable to detect patient positioning errors, and the R-value has good sensitivity to phantom displacements. A gamma comparison analysis allows one to determine alert thresholds to detect phantom shifts or relatively large rotations. The combined use of the IQM and SoftDiso allows for fast identification of both delivery and setup errors and substantially reduces the impact of error identification and correction on the treatment workflow.

Combined use of a transmission detector and an epid-based in vivo dose monitoring system in external beam whole breast irradiation: A study with an anthropomorphic female phantom / Arilli C.; Wandael Y.; Galeotti C.; Marrazzo L.; Calusi S.; Grusio M.; Desideri I.; Fusi F.; Piermattei A.; Pallotta S.; Talamonti C.. - In: APPLIED SCIENCES. - ISSN 2076-3417. - ELETTRONICO. - 10:(2020), pp. 1-12. [10.3390/app10217611]

Combined use of a transmission detector and an epid-based in vivo dose monitoring system in external beam whole breast irradiation: A study with an anthropomorphic female phantom

Arilli C.;Wandael Y.;Galeotti C.;Marrazzo L.;Calusi S.;Desideri I.;Fusi F.;Pallotta S.;Talamonti C.
2020

Abstract

We evaluate the combined usage of two systems, the Integral Quality Monitor (IQM) transmission detector and SoftDiso software, for in vivo dose monitoring by simultaneous detection of delivery and patient setup errors in whole breast irradiation. An Alderson RANDO phantom was adapted with silicon breast prostheses to mimic the female anatomy. Plans with simulated delivery errors were created from a reference left breast plan, and patient setup errors were simulated by moving the phantom. Deviations from reference values recorded by both monitoring systems were measured for all plans and phantom positions. A 2D global gamma analysis was performed in SoftDiso for all phantom displacements. Both IQM signals and SoftDiso R-values are sensitive to small MU variations. However, only IQM is sensitive to jaw position variations. Conversely, IQM is unable to detect patient positioning errors, and the R-value has good sensitivity to phantom displacements. A gamma comparison analysis allows one to determine alert thresholds to detect phantom shifts or relatively large rotations. The combined use of the IQM and SoftDiso allows for fast identification of both delivery and setup errors and substantially reduces the impact of error identification and correction on the treatment workflow.
2020
10
1
12
Goal 3: Good health and well-being for people
Arilli C.; Wandael Y.; Galeotti C.; Marrazzo L.; Calusi S.; Grusio M.; Desideri I.; Fusi F.; Piermattei A.; Pallotta S.; Talamonti C.
File in questo prodotto:
File Dimensione Formato  
applsci-10-07611-v2.pdf

accesso aperto

Tipologia: Pdf editoriale (Version of record)
Licenza: Open Access
Dimensione 1.19 MB
Formato Adobe PDF
1.19 MB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1215492
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact