Periodic dose assessment is quintessential for dynamic dose adjustment and quality control of continuous renal replacement therapy (CRRT) in critically ill patients with acute kidney injury (AKI). The flows-based methods to estimate dose are easy and reproducible methods to quantify (estimate)CRRT dose at the bedside. In particular, quantification of effluent flow and, mainly, the current dose (adjusted for dialysate, replacement, blood flows and net ultrafiltration) is routinely used in clinical practice. Unfortunately, these methods are critically influenced by several external unpredictable factors; the estimated dose often overestimates the real biological delivered dose quantified through measurement of urea clearance (the current effective delivered dose). Although the current effective delivered dose is undoubtedly more precise than the flows-based dose estimation in quantifying CRRT efficacy, some limitations are reported for the urea-based measurement of dose. This paper aims to describe the standard of practice for dose quantification in critically ill patients with AKI undergoing CRRT in the intensive care unit. Pitfalls of currents methods will be underlined, along with solutions potentially applicable to obtain more precise results in terms of 1) adequate marker solutes that should be used in accordance with the clinical scenario, 2) correct sampling procedures depending on the chosen indicator of transmembrane removal 3) formulas for calculations 4) quality controls and benchmark indicators.

Methods for dose quantification in continuous renal replacement therapy: towards a more precise approach / Villa, Gianluca; Fabbri, Sergio; Samoni, Sara; Cecchi, Matteo; Fioccola, Antonio; Scirè-Calabrisotto, Caterina; Mari, Gaia; Pomarè Montin, Diego; Romagnoli, Stefano. - In: ARTIFICIAL ORGANS. - ISSN 0160-564X. - ELETTRONICO. - (2021), pp. 1-10. [10.1111/aor.13991]

Methods for dose quantification in continuous renal replacement therapy: towards a more precise approach

Villa, Gianluca
;
Fabbri, Sergio;Samoni, Sara;Cecchi, Matteo;Fioccola, Antonio;Scirè-Calabrisotto, Caterina;Mari, Gaia;Pomarè Montin, Diego;Romagnoli, Stefano
2021

Abstract

Periodic dose assessment is quintessential for dynamic dose adjustment and quality control of continuous renal replacement therapy (CRRT) in critically ill patients with acute kidney injury (AKI). The flows-based methods to estimate dose are easy and reproducible methods to quantify (estimate)CRRT dose at the bedside. In particular, quantification of effluent flow and, mainly, the current dose (adjusted for dialysate, replacement, blood flows and net ultrafiltration) is routinely used in clinical practice. Unfortunately, these methods are critically influenced by several external unpredictable factors; the estimated dose often overestimates the real biological delivered dose quantified through measurement of urea clearance (the current effective delivered dose). Although the current effective delivered dose is undoubtedly more precise than the flows-based dose estimation in quantifying CRRT efficacy, some limitations are reported for the urea-based measurement of dose. This paper aims to describe the standard of practice for dose quantification in critically ill patients with AKI undergoing CRRT in the intensive care unit. Pitfalls of currents methods will be underlined, along with solutions potentially applicable to obtain more precise results in terms of 1) adequate marker solutes that should be used in accordance with the clinical scenario, 2) correct sampling procedures depending on the chosen indicator of transmembrane removal 3) formulas for calculations 4) quality controls and benchmark indicators.
2021
1
10
Villa, Gianluca; Fabbri, Sergio; Samoni, Sara; Cecchi, Matteo; Fioccola, Antonio; Scirè-Calabrisotto, Caterina; Mari, Gaia; Pomarè Montin, Diego; Romagnoli, Stefano
File in questo prodotto:
File Dimensione Formato  
Artificial Organs - 2021 - Villa - Methods for dose quantification in continuous renal replacement therapy Toward a more.pdf

accesso aperto

Tipologia: Pdf editoriale (Version of record)
Licenza: Creative commons
Dimensione 596.69 kB
Formato Adobe PDF
596.69 kB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1237695
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact