The aim of this study was to develop and apply an evaluation method for assessing the accuracy of a novel 3D EPID back-projection algorithm for in vivo dosimetry. The novel algorithm of Dosimetry Check (DC) 5.8 was evaluated. A slab phantom homogeneously filled, or with air and bone inserts, was used for fluence reconstruction of different squared fields. VMAT plans in different anatomical sites were delivered on an anthropomorphic phantom. Dose distributions were measured with radiochromic films. The 2D Gamma Agreement Index (GAI) between the DC and the film dose distributions (3%, 3 mm) was computed for assessing the accuracy of the algorithm. GAIs between films and TPS and between DC and TPS were also computed. The fluence reconstruction accuracy was within 2% for all squared fields in the three slabs’ configurations. The GAI between the DC and the film was 92.7% in the prostate, 92.9% in the lung, 96.6% in the head and the neck, and 94.6% in the brain. An evaluation method for assessing the accuracy of a novel EPID algorithm was developed. The DC algorithm was shown to be able to accurately reconstruct doses in all anatomic sites, including the lung. The methodology described in the present study can be applied to any EPID back-projection in vivo algorithm.

A validation method for EPID in vivo dosimetry algorithms / Esposito M.; Marrazzo L.; Vanzi E.; Russo S.; Pallotta S.; Talamonti C.. - In: APPLIED SCIENCES. - ISSN 2076-3417. - ELETTRONICO. - 11:(2021), pp. 10715-10725. [10.3390/app112210715]

A validation method for EPID in vivo dosimetry algorithms

Marrazzo L.;Russo S.;Pallotta S.;Talamonti C.
2021

Abstract

The aim of this study was to develop and apply an evaluation method for assessing the accuracy of a novel 3D EPID back-projection algorithm for in vivo dosimetry. The novel algorithm of Dosimetry Check (DC) 5.8 was evaluated. A slab phantom homogeneously filled, or with air and bone inserts, was used for fluence reconstruction of different squared fields. VMAT plans in different anatomical sites were delivered on an anthropomorphic phantom. Dose distributions were measured with radiochromic films. The 2D Gamma Agreement Index (GAI) between the DC and the film dose distributions (3%, 3 mm) was computed for assessing the accuracy of the algorithm. GAIs between films and TPS and between DC and TPS were also computed. The fluence reconstruction accuracy was within 2% for all squared fields in the three slabs’ configurations. The GAI between the DC and the film was 92.7% in the prostate, 92.9% in the lung, 96.6% in the head and the neck, and 94.6% in the brain. An evaluation method for assessing the accuracy of a novel EPID algorithm was developed. The DC algorithm was shown to be able to accurately reconstruct doses in all anatomic sites, including the lung. The methodology described in the present study can be applied to any EPID back-projection in vivo algorithm.
2021
11
10715
10725
Esposito M.; Marrazzo L.; Vanzi E.; Russo S.; Pallotta S.; Talamonti C.
File in questo prodotto:
File Dimensione Formato  
A Validation Method for EPID In Vivo Dosimetry Algorithms | Enhanced Reader.pdf

accesso aperto

Tipologia: Pdf editoriale (Version of record)
Licenza: Creative commons
Dimensione 20.68 MB
Formato Adobe PDF
20.68 MB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1250416
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
social impact