The 4D GRAPH-X (Dynamic GRAting-based PHase contrast x-ray imaging) project aims at developing a prototype of an x-ray grating-based phase-contrast imaging scanner in a laboratory setting, which is based on the Moirè single-shot acquisition method in order to be optimized for analysing moving objects (in the specific case, a dynamic thorax phantom), that could evolve into a suitable tool for biomedical applications although it can be extended to other application fields. When designing an x-ray Talbot-Lau interferometer, high visibility and sensitivity are two important figures of merit, strictly related to the performance of the system in obtaining high quality phase contrast and dark-field images. Wave field simulations are performed to optimize the setup specifications and construct a high-resolution and high-sensitivity imaging system. In this work, the design of a dynamic imaging setup using a conventional milli-focus x-ray source is presented. Optimization by wave front simulations leads to a symmetric configuration with 5.25 μm pitch at third Talbot order and 45 keV design energy. The simulated visibility is about 22%. Results from GATE based Monte Carlo simulations show a 19% transmission percentage of the incoming beam into the detector after passing through all the gratings and the sample. Such results are promising in view of building a system optimized for dynamic imaging.
X-ray grating interferometry design for the 4D GRAPH-X system / Patera A.; Arboleda C.; Ferrero V.; Fiorina E.; Jefimovs K.; Giudice A.L.; Milian F.M.; Mereu P.; Pallotta S.; Ramello L.; Re A.; Visca L.; Cerello P.. - In: JOURNAL OF PHYSICS D. APPLIED PHYSICS. - ISSN 0022-3727. - STAMPA. - 55:(2022), pp. 045103-045103. [10.1088/1361-6463/ac2fd6]
X-ray grating interferometry design for the 4D GRAPH-X system
Patera A.;Pallotta S.;
2022
Abstract
The 4D GRAPH-X (Dynamic GRAting-based PHase contrast x-ray imaging) project aims at developing a prototype of an x-ray grating-based phase-contrast imaging scanner in a laboratory setting, which is based on the Moirè single-shot acquisition method in order to be optimized for analysing moving objects (in the specific case, a dynamic thorax phantom), that could evolve into a suitable tool for biomedical applications although it can be extended to other application fields. When designing an x-ray Talbot-Lau interferometer, high visibility and sensitivity are two important figures of merit, strictly related to the performance of the system in obtaining high quality phase contrast and dark-field images. Wave field simulations are performed to optimize the setup specifications and construct a high-resolution and high-sensitivity imaging system. In this work, the design of a dynamic imaging setup using a conventional milli-focus x-ray source is presented. Optimization by wave front simulations leads to a symmetric configuration with 5.25 μm pitch at third Talbot order and 45 keV design energy. The simulated visibility is about 22%. Results from GATE based Monte Carlo simulations show a 19% transmission percentage of the incoming beam into the detector after passing through all the gratings and the sample. Such results are promising in view of building a system optimized for dynamic imaging.File | Dimensione | Formato | |
---|---|---|---|
Patera_2022_J._Phys._D__Appl._Phys._55_045103 X-ray grating interferometry design for the 4D GRAPH-X system.pdf
accesso aperto
Tipologia:
Pdf editoriale (Version of record)
Licenza:
Tutti i diritti riservati
Dimensione
1.32 MB
Formato
Adobe PDF
|
1.32 MB | Adobe PDF |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.