Objective. This paper describes the procedure to calibrate the three-dimensional (3D) proton stopping power relative to water ( SPR) maps measured by the proton computed tomography (pCT) apparatus of the Istituto Nazionale di Fisica Nucleare (INFN, Italy). Measurements performed on water phantoms are used to validate the method. The calibration allowed for achieving measurement accuracy and reproducibility to levels below 1%. Approach. The INFN pCT system is made of a silicon tracker for proton trajectory determination followed by a YAG:Ce calorimeter for energy measurement. To perform the calibration, the apparatus has been exposed to protons of energies ranging from 83 to 210 MeV. Using the tracker, a position-dependent calibration has been implemented to keep the energy response uniform across the calorimeter. Moreover, correction algorithms have been developed to reconstruct the proton energy when this is shared in more than one crystal and to consider the energy loss in the non-uniform apparatus material. To verify the calibration and its reproducibility, water phantoms have been imaged with the pCT system during two data-taking sessions. Main results. The energy resolution of the pCT calorimeter resulted to be sigma(E)/E congruent to 0.9% at 196.5 MeV. The average values of the water SPR in fiducial volumes of the control phantoms have been calculated to be 0.995 +/- 0.002. The image non-uniformities were below 1%. No appreciable variation of the SPR and uniformity values between the two data-taking sessions could be identified. Significance. This work demonstrates the accuracy and reproducibility of the calibration of the INFN pCT system at a level below 1%. Moreover, the uniformity of the energy response keeps the image artifacts at a low level even in the presence of calorimeter segmentation and tracker material nonuniformities. The implemented calibration technique allows the INFN-pCT system to face applications where the precision of the SPR 3D maps is of paramount importance.

The INFN proton computed tomography system for relative stopping power measurements: calibration and verification / Scaringella, Monica; Bruzzi, Mara; Farace, Paolo; Fogazzi, Elena; Righetto, Roberto; Rit, Simon; Tommasino, Francesco; Verroi, Enrico; Civinini, Carlo. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - ELETTRONICO. - 68:(2023), pp. 154001-154020. [10.1088/1361-6560/ace2a8]

The INFN proton computed tomography system for relative stopping power measurements: calibration and verification

Scaringella, Monica
;
Bruzzi, Mara;Civinini, Carlo
2023

Abstract

Objective. This paper describes the procedure to calibrate the three-dimensional (3D) proton stopping power relative to water ( SPR) maps measured by the proton computed tomography (pCT) apparatus of the Istituto Nazionale di Fisica Nucleare (INFN, Italy). Measurements performed on water phantoms are used to validate the method. The calibration allowed for achieving measurement accuracy and reproducibility to levels below 1%. Approach. The INFN pCT system is made of a silicon tracker for proton trajectory determination followed by a YAG:Ce calorimeter for energy measurement. To perform the calibration, the apparatus has been exposed to protons of energies ranging from 83 to 210 MeV. Using the tracker, a position-dependent calibration has been implemented to keep the energy response uniform across the calorimeter. Moreover, correction algorithms have been developed to reconstruct the proton energy when this is shared in more than one crystal and to consider the energy loss in the non-uniform apparatus material. To verify the calibration and its reproducibility, water phantoms have been imaged with the pCT system during two data-taking sessions. Main results. The energy resolution of the pCT calorimeter resulted to be sigma(E)/E congruent to 0.9% at 196.5 MeV. The average values of the water SPR in fiducial volumes of the control phantoms have been calculated to be 0.995 +/- 0.002. The image non-uniformities were below 1%. No appreciable variation of the SPR and uniformity values between the two data-taking sessions could be identified. Significance. This work demonstrates the accuracy and reproducibility of the calibration of the INFN pCT system at a level below 1%. Moreover, the uniformity of the energy response keeps the image artifacts at a low level even in the presence of calorimeter segmentation and tracker material nonuniformities. The implemented calibration technique allows the INFN-pCT system to face applications where the precision of the SPR 3D maps is of paramount importance.
2023
68
154001
154020
Goal 3: Good health and well-being
Scaringella, Monica; Bruzzi, Mara; Farace, Paolo; Fogazzi, Elena; Righetto, Roberto; Rit, Simon; Tommasino, Francesco; Verroi, Enrico; Civinini, Carlo
File in questo prodotto:
File Dimensione Formato  
Scaringella_2023_Phys._Med._Biol._68_154001 (1).pdf

accesso aperto

Tipologia: Pdf editoriale (Version of record)
Licenza: Open Access
Dimensione 2.07 MB
Formato Adobe PDF
2.07 MB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1329911
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact