PurposeThe lack of a reliable and reproducible technique to ensure a constantly low temperature of the graft during kidney transplantation (KT) may be a cause of renal nonfunction. The aim of this review was to assess all the methods and devices available to ensure hypothermia during vascular anastomosis in KT.MethodsA literature search was conducted through May 2022 using PubMed/Medline, Cochrane Library, Embase and Web of Science databases. Preferred Reporting Items for Systematic Reviews and Meta-analysis (PRISMA) guidelines were followed to identify eligible studies. The review protocol was registered on PROSPERO (CRD42022326550).ResultsA total of 20 studies reporting on four hypothermia techniques met our inclusion criteria. Simple instillation of cold serum is not sufficient, the graft reaching up to 33 celcius at the end of warm ischemia time (WIT). Plastic bags filled with ice slush have questionable efficiency. The use of a gauze jacket filled with ice-slush was reported in 12/20 studies. It ensures a graft temperature up to 20.3 celcius at the end of WIT. Some concerns have been linked to potentially inhomogeneous parenchymal cooling and secondary ileus. Novel devices with continuous flow of ice-cold solution around the graft might overcome these limitations, showing a renal temperature below 20 celcius at all times during KT.ConclusionThe gauze filled with ice slush is the most common technique, but several aspects can be improved. Novel devices in the form of cold-ischemia jackets can ensure a lower and more stable temperature of the graft during KT, leading to higher efficiency and reproducibility.
Systematic review of techniques and devices used to avoid warm ischemia time injury during kidney transplantation / Andras, Iulia; Piana, Alberto; Verri, Paolo; Telecan, Teodora; Gallioli, Andrea; Prudhomme, Thomas; Hevia, Vital; Baboudjian, Michael; Boissier, Romain; Crisan, Nicolae; Campi, Riccardo; Breda, Alberto; Territo, Angelo. - In: WORLD JOURNAL OF UROLOGY. - ISSN 0724-4983. - ELETTRONICO. - 41:(2023), pp. 993-1003. [10.1007/s00345-023-04328-9]
Systematic review of techniques and devices used to avoid warm ischemia time injury during kidney transplantation
Campi, Riccardo;
2023
Abstract
PurposeThe lack of a reliable and reproducible technique to ensure a constantly low temperature of the graft during kidney transplantation (KT) may be a cause of renal nonfunction. The aim of this review was to assess all the methods and devices available to ensure hypothermia during vascular anastomosis in KT.MethodsA literature search was conducted through May 2022 using PubMed/Medline, Cochrane Library, Embase and Web of Science databases. Preferred Reporting Items for Systematic Reviews and Meta-analysis (PRISMA) guidelines were followed to identify eligible studies. The review protocol was registered on PROSPERO (CRD42022326550).ResultsA total of 20 studies reporting on four hypothermia techniques met our inclusion criteria. Simple instillation of cold serum is not sufficient, the graft reaching up to 33 celcius at the end of warm ischemia time (WIT). Plastic bags filled with ice slush have questionable efficiency. The use of a gauze jacket filled with ice-slush was reported in 12/20 studies. It ensures a graft temperature up to 20.3 celcius at the end of WIT. Some concerns have been linked to potentially inhomogeneous parenchymal cooling and secondary ileus. Novel devices with continuous flow of ice-cold solution around the graft might overcome these limitations, showing a renal temperature below 20 celcius at all times during KT.ConclusionThe gauze filled with ice slush is the most common technique, but several aspects can be improved. Novel devices in the form of cold-ischemia jackets can ensure a lower and more stable temperature of the graft during KT, leading to higher efficiency and reproducibility.| File | Dimensione | Formato | |
|---|---|---|---|
|
s00345-023-04328-9.pdf
Accesso chiuso
Tipologia:
Pdf editoriale (Version of record)
Licenza:
Tutti i diritti riservati
Dimensione
1.05 MB
Formato
Adobe PDF
|
1.05 MB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.



