: We aimed to describe a variation of the surgical technique for the ab interno implantation of the XEN Gel Stent, which, in our experience, is yielding very successful results. The injection of 0.1 mL of air and then of 0.1 mL of a dispersive viscoelastic into the subconjunctival space at the beginning of the surgery allows one to perform a mechanical dissection between the conjunctiva and the Tenon's capsule, creating a real space. In total, 20 eyes of 16 patients underwent the implantation of a stent gel through the "Air and Visco" technique. We retrospectively analyzed the results. We obtained a reduction in the IOP from an average of 18.3 ± 2.2 mmHg preoperatively to at 13.5 ± 3.5 mmHg at month 12. The needling rate was 20%. We did not register any cases of hypotony (IOP < 6 mmHg), hypotony maculopathy or choroidal detachment. The "Air and Visco" technique allows one to correctly place the device in the subconjunctival space, which the pneumo- and visco-dissection transforms into a real space. This enables an easier surgical performance and more predictable postoperative results, with a low needling rate and reintervention in the follow-up period. It also ensures a greater safety profile because the presence of the OVD on the bleb prevents a sudden lowering of the IOP, eliminating complications such as hypotony, hypotony maculopathy and choroidal detachment in our cohort.

“Air and Visco” Technique: A Promising Innovation in the Surgical Implantation of the Xen Gel Stent Device / Franco, Fabrizio; Serino, Federica; Giansanti, Fabrizio. - In: VISION. - ISSN 2411-5150. - ELETTRONICO. - 7:(2023), pp. 71.0-71.0. [10.3390/vision7040071]

“Air and Visco” Technique: A Promising Innovation in the Surgical Implantation of the Xen Gel Stent Device

Franco, Fabrizio;Serino, Federica;Giansanti, Fabrizio
2023

Abstract

: We aimed to describe a variation of the surgical technique for the ab interno implantation of the XEN Gel Stent, which, in our experience, is yielding very successful results. The injection of 0.1 mL of air and then of 0.1 mL of a dispersive viscoelastic into the subconjunctival space at the beginning of the surgery allows one to perform a mechanical dissection between the conjunctiva and the Tenon's capsule, creating a real space. In total, 20 eyes of 16 patients underwent the implantation of a stent gel through the "Air and Visco" technique. We retrospectively analyzed the results. We obtained a reduction in the IOP from an average of 18.3 ± 2.2 mmHg preoperatively to at 13.5 ± 3.5 mmHg at month 12. The needling rate was 20%. We did not register any cases of hypotony (IOP < 6 mmHg), hypotony maculopathy or choroidal detachment. The "Air and Visco" technique allows one to correctly place the device in the subconjunctival space, which the pneumo- and visco-dissection transforms into a real space. This enables an easier surgical performance and more predictable postoperative results, with a low needling rate and reintervention in the follow-up period. It also ensures a greater safety profile because the presence of the OVD on the bleb prevents a sudden lowering of the IOP, eliminating complications such as hypotony, hypotony maculopathy and choroidal detachment in our cohort.
2023
7
0
0
Franco, Fabrizio; Serino, Federica; Giansanti, Fabrizio
File in questo prodotto:
File Dimensione Formato  
vision-07-00071.pdf

accesso aperto

Tipologia: Pdf editoriale (Version of record)
Licenza: Creative commons
Dimensione 427.69 kB
Formato Adobe PDF
427.69 kB Adobe PDF