Objective: This study aimed to compare two commercially available off the shelf branched endografts for thoraco-abdominal aortic aneurysm (TAAA) repair, namely the E-nside (Artivion) and Zenith t-Branch (Cook Medical) devices. Methods: This multicentre retrospective study (2020 - 2023) included patients treated by branched endovascular aortic repair (BEVAR) for TAAA using the inner branched E-nside or the outer branched t-Branch. Endpoints were 30 day technical success and major adverse events (MAEs) as well as 1 year freedom from target vessel instability and main endograft instability. Results: The study included 163 patients: 79 (307 target vessels) treated with E-nside and 84 (325 target vessels) with t-branch. Aneurysm extent was I - III in 91 patients (55.8%; 47% of E-nside and 66% of t-Branch) and IV in 72 patients (44.2%; 53% of E-nside and 34% of t-Branch) (p = .011). An adjunctive proximal thoracic endograft was used in 43% of E-nside vs. 69% of t-Branch (p < .001), with less frequent thoracic endografting (14% vs. 76%; p < .001) and shorter length of coverage (p = .024) in extent IV TAAA treated by E-nside. E-nside cases had shorter bridging length of renal arteries (66 ± 17 mm vs. 76 ± 20 mm; p < .010) and less frequent use of a distal bifurcated endograft (53% vs. 80%; p < .001). Comparing 30 day results, mortality was 1% vs. 2% (p = .62), any MAE occurred in 18% vs. 21% (p = .55), the stroke rate was 3% vs. 0% (p = .23), and the elective spinal cord ischaemia rate was 5% vs. 8% (p = .40) for E-nside and t-Branch, respectively. At 1 year, freedom from target vessel instability was 96 ± 3% for E-nside and 95 ± 3% for t-Branch (p = .58), and freedom from endograft instability was 98 ± 2% vs. 97 ± 3% (p = .46) respectively. Conclusion: Both off the shelf devices provided excellent early and 1 year results. The E-nside may require shorter thoracic aorta coverage and bridging length for the renal arteries, and less frequent implantation of concomitant proximal thoracic or distal abdominal bifurcated endograft. However, these aspects did not determine significant differences in clinical outcomes.

Outcomes of Off the Shelf Outer Branched Versus Inner Branched Endografts in the Treatment of Thoraco-abdominal Aortic Aneurysm in the B.R.I.O. (BRanched Inner - Outer) Study Group / Andrea Spertino, Franco Grego, Sara Di Gregorio, Massimo Lenti, Giacomo Isernia, Ciro Ferrer, Gabriele Piffaretti, Francesca Mauri, Luca Mezzetto, Davide MastrorillI, Roberto Silingardi, Nicola Leone, Yamume Tshomba, Tommaso Donati, Simona Sica, Giovanni Tinelli, Raffaele Pulli, Aaron Fargion, Carlo Pratesi, Luca Di Marzo, Alessia Di Girolamo, Assunta Orticelli, Stefano Mechelagnoli, Emiliano Chisci, Stefano Fazzini, Fabio Massimo Oddi, Arnaldo Ippoliti. - In: EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY. - ISSN 1078-5884. - STAMPA. - (In corso di stampa), pp. 0-0.

Outcomes of Off the Shelf Outer Branched Versus Inner Branched Endografts in the Treatment of Thoraco-abdominal Aortic Aneurysm in the B.R.I.O. (BRanched Inner - Outer) Study Group.

Raffaele Pulli;Aaron Fargion;Carlo Pratesi;
In corso di stampa

Abstract

Objective: This study aimed to compare two commercially available off the shelf branched endografts for thoraco-abdominal aortic aneurysm (TAAA) repair, namely the E-nside (Artivion) and Zenith t-Branch (Cook Medical) devices. Methods: This multicentre retrospective study (2020 - 2023) included patients treated by branched endovascular aortic repair (BEVAR) for TAAA using the inner branched E-nside or the outer branched t-Branch. Endpoints were 30 day technical success and major adverse events (MAEs) as well as 1 year freedom from target vessel instability and main endograft instability. Results: The study included 163 patients: 79 (307 target vessels) treated with E-nside and 84 (325 target vessels) with t-branch. Aneurysm extent was I - III in 91 patients (55.8%; 47% of E-nside and 66% of t-Branch) and IV in 72 patients (44.2%; 53% of E-nside and 34% of t-Branch) (p = .011). An adjunctive proximal thoracic endograft was used in 43% of E-nside vs. 69% of t-Branch (p < .001), with less frequent thoracic endografting (14% vs. 76%; p < .001) and shorter length of coverage (p = .024) in extent IV TAAA treated by E-nside. E-nside cases had shorter bridging length of renal arteries (66 ± 17 mm vs. 76 ± 20 mm; p < .010) and less frequent use of a distal bifurcated endograft (53% vs. 80%; p < .001). Comparing 30 day results, mortality was 1% vs. 2% (p = .62), any MAE occurred in 18% vs. 21% (p = .55), the stroke rate was 3% vs. 0% (p = .23), and the elective spinal cord ischaemia rate was 5% vs. 8% (p = .40) for E-nside and t-Branch, respectively. At 1 year, freedom from target vessel instability was 96 ± 3% for E-nside and 95 ± 3% for t-Branch (p = .58), and freedom from endograft instability was 98 ± 2% vs. 97 ± 3% (p = .46) respectively. Conclusion: Both off the shelf devices provided excellent early and 1 year results. The E-nside may require shorter thoracic aorta coverage and bridging length for the renal arteries, and less frequent implantation of concomitant proximal thoracic or distal abdominal bifurcated endograft. However, these aspects did not determine significant differences in clinical outcomes.
In corso di stampa
0
0
Andrea Spertino, Franco Grego, Sara Di Gregorio, Massimo Lenti, Giacomo Isernia, Ciro Ferrer, Gabriele Piffaretti, Francesca Mauri, Luca Mezzetto, Dav...espandi
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1357568
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact