Detection of SARS-CoV-2 in bronchoalveolar lavage (BAL) is considered as a promising alternative method to detect COVID-19 infection. STANDARDTM TM M10 SARS-CoV-2 assay on 150 negative and 50 positives BAL samples for SARS-CoV-2 showed 96% sensitivity, 100% specificity compared to AllplexTM TM SARS-CoV-2 assay and a 31.25 genomic copies/mL limit of detection.
Evaluation of STANDARD™ M10 SARS-CoV-2 from bronchoalveolar lavage samples / Bartolini, Andrea; Morecchiato, Fabio; Antonelli, Alberto; Malentacchi, Francesca; Rossolini, Gian Maria; Pollini, Simona. - In: DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE. - ISSN 0732-8893. - ELETTRONICO. - 110:(2024), pp. 116466.0-116466.0. [10.1016/j.diagmicrobio.2024.116466]
Evaluation of STANDARD™ M10 SARS-CoV-2 from bronchoalveolar lavage samples
Morecchiato, Fabio;Antonelli, Alberto;Malentacchi, Francesca;Rossolini, Gian Maria;Pollini, Simona
2024
Abstract
Detection of SARS-CoV-2 in bronchoalveolar lavage (BAL) is considered as a promising alternative method to detect COVID-19 infection. STANDARDTM TM M10 SARS-CoV-2 assay on 150 negative and 50 positives BAL samples for SARS-CoV-2 showed 96% sensitivity, 100% specificity compared to AllplexTM TM SARS-CoV-2 assay and a 31.25 genomic copies/mL limit of detection.File | Dimensione | Formato | |
---|---|---|---|