Microsystems represent an alternative but proficient approach of analysis outside the laboratory, and their use could help in reducing the impact of pre-analytical errors, in particular in challenging newborn samples. The study purpose is to compare the Horiba Microsemi CRP LC-767G system for rapid 3-part complete blood count (CBC) and C-reactive protein (CRP) determination with the laboratory reference systems (respectively Sysmex XN-9100 (TM) and Roche Cobas (R) c702) in samples of adult patients and newborns hospitalized in the neonatal intensive care unit (NICU) samples. The comparison between the analyzers was performed through Passing-Bablok regression analysis and Bland-Altman plot. One hundred eighty-three blood samples were analyzed. The regression analysis results, performed in the newborn (n = 70) and in adult (n = 113) populations, showed a good agreement between the instruments. The evaluation of the Bland-Altman plots showed comparable values of bias < 10% for most of the parameters, but not for MPV, lymphocyte, and monocyte count. Conclusion: The comparison between the Microsemi CRP LC-767G system and the laboratory instrumentations demonstrated comparable results. The Microsemi CRP LC-767G system provides reliable analytical data and faster turnaround time, particularly useful in NICU.

Rapid complete blood count and C-reactive protein determination with the Horiba Microsemi analyzer: the experience in neonatal intensive care unit of Careggi University Hospital / Nencini, Francesca; Bonari, Alessandro; Romano, Francesca; Ciullini Mannurita, Sara; Mongia, Alessandra; Garieri, Maria; Dani, Carlo; Manzo, Orazio; Pontieri, Maria; Fanelli, Alessandra. - In: EUROPEAN JOURNAL OF PEDIATRICS. - ISSN 0340-6199. - STAMPA. - 183:(2024), pp. 4477-4490. [10.1007/s00431-024-05695-0]

Rapid complete blood count and C-reactive protein determination with the Horiba Microsemi analyzer: the experience in neonatal intensive care unit of Careggi University Hospital

Dani, Carlo;
2024

Abstract

Microsystems represent an alternative but proficient approach of analysis outside the laboratory, and their use could help in reducing the impact of pre-analytical errors, in particular in challenging newborn samples. The study purpose is to compare the Horiba Microsemi CRP LC-767G system for rapid 3-part complete blood count (CBC) and C-reactive protein (CRP) determination with the laboratory reference systems (respectively Sysmex XN-9100 (TM) and Roche Cobas (R) c702) in samples of adult patients and newborns hospitalized in the neonatal intensive care unit (NICU) samples. The comparison between the analyzers was performed through Passing-Bablok regression analysis and Bland-Altman plot. One hundred eighty-three blood samples were analyzed. The regression analysis results, performed in the newborn (n = 70) and in adult (n = 113) populations, showed a good agreement between the instruments. The evaluation of the Bland-Altman plots showed comparable values of bias < 10% for most of the parameters, but not for MPV, lymphocyte, and monocyte count. Conclusion: The comparison between the Microsemi CRP LC-767G system and the laboratory instrumentations demonstrated comparable results. The Microsemi CRP LC-767G system provides reliable analytical data and faster turnaround time, particularly useful in NICU.
2024
183
4477
4490
Goal 3: Good health and well-being
Nencini, Francesca; Bonari, Alessandro; Romano, Francesca; Ciullini Mannurita, Sara; Mongia, Alessandra; Garieri, Maria; Dani, Carlo; Manzo, Orazio; P...espandi
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1401933
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact