The Tendyne valve (Abbott Structural) is a self-expanding, nitinol prothesis with a double-frame design that contains a trileaflet porcine pericardial valve anatomically shaped for the mitral annulus. The prosthesis, which is delivered via a 34Fr transapical sheath, is uniquely anchored with a tether connected to an epicardial hemostatic pad at the apex. The Tendyne prosthesis is retrievable and repositionable after full deployment. Delayed atrial dislocation of Tendyne causing hemolysis and congestive heart failure symptoms has been recently described.1 This condition has occurred in a small number of patients, and there are anecdotal reports of successful retensioning of the apical tether a few weeks after the initial procedure.2 We report a case of late dislocation of a transcatheter mitral prosthesis followed by unsuccessful retensioning and open surgery.

Late displacement and ineffective retensioning of a transapical transcatheter mitral prosthesis / Cerillo, Alfredo Giuseppe; Paduvakis, Jorgos; Petrini, Francesca; Stefano, Pierluigi. - In: JTCVS TECHNIQUES. - ISSN 2666-2507. - STAMPA. - 20:(2023), pp. 55-57. [10.1016/j.xjtc.2023.05.018]

Late displacement and ineffective retensioning of a transapical transcatheter mitral prosthesis

Cerillo, Alfredo Giuseppe;Petrini, Francesca;Stefano, Pierluigi
2023

Abstract

The Tendyne valve (Abbott Structural) is a self-expanding, nitinol prothesis with a double-frame design that contains a trileaflet porcine pericardial valve anatomically shaped for the mitral annulus. The prosthesis, which is delivered via a 34Fr transapical sheath, is uniquely anchored with a tether connected to an epicardial hemostatic pad at the apex. The Tendyne prosthesis is retrievable and repositionable after full deployment. Delayed atrial dislocation of Tendyne causing hemolysis and congestive heart failure symptoms has been recently described.1 This condition has occurred in a small number of patients, and there are anecdotal reports of successful retensioning of the apical tether a few weeks after the initial procedure.2 We report a case of late dislocation of a transcatheter mitral prosthesis followed by unsuccessful retensioning and open surgery.
2023
20
55
57
Goal 3: Good health and well-being
Cerillo, Alfredo Giuseppe; Paduvakis, Jorgos; Petrini, Francesca; Stefano, Pierluigi
File in questo prodotto:
File Dimensione Formato  
PIIS2666250723001931.pdf

accesso aperto

Tipologia: Pdf editoriale (Version of record)
Licenza: Open Access
Dimensione 913.11 kB
Formato Adobe PDF
913.11 kB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1406559
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact