Purpose: This study presents a detailed optical characterization of three intraocular lenses (IOLs) comparing two so called “enhanced range of field”, Evolux IOL and Tecnis Eyhance IOL and one so called “narrow range of field”, a standard monofocal IOL Acrysof IQ. Methods: The measurements are performed using the PMTF optical bench, basing on the Modulation Transfer Function (MTF) to evaluate the optical performance of each lens. The MTFa, representing the area under the MTF curve, is utilized as a key and synthetic metric to quantify performance across different spatial frequencies, providing insights into the IOLs’ behavior as defocus varies. Results: The results highlight the strengths and the weaknesses of the IOLs, with the Evolux showing a broadest depth of focus. The USAF resolution target is used to qualitatively assess the image reproduction at far, intermediate, and near distances, offering a visual representation of the IOLs’ capabilities. Conclusions: These findings provide valuable information for selecting IOLs based on patient-specific visual requirements in cataract and refractive surgery.
Comparison of optical performances of intraocular lenses with different depth of field / Tommasi F.; Giorgetti A.; Monnetti C.; Fini L.; Romualdi G.; Mencucci R.; Gurioli M.; Cavalieri S.. - In: JOURNAL OF OPTOMETRY. - ISSN 1888-4296. - ELETTRONICO. - 18:(2025), pp. 100546.0-100546.0. [10.1016/j.optom.2025.100546]
Comparison of optical performances of intraocular lenses with different depth of field
Tommasi F.
;Giorgetti A.;Fini L.;Romualdi G.;Mencucci R.;Gurioli M.;Cavalieri S.
2025
Abstract
Purpose: This study presents a detailed optical characterization of three intraocular lenses (IOLs) comparing two so called “enhanced range of field”, Evolux IOL and Tecnis Eyhance IOL and one so called “narrow range of field”, a standard monofocal IOL Acrysof IQ. Methods: The measurements are performed using the PMTF optical bench, basing on the Modulation Transfer Function (MTF) to evaluate the optical performance of each lens. The MTFa, representing the area under the MTF curve, is utilized as a key and synthetic metric to quantify performance across different spatial frequencies, providing insights into the IOLs’ behavior as defocus varies. Results: The results highlight the strengths and the weaknesses of the IOLs, with the Evolux showing a broadest depth of focus. The USAF resolution target is used to qualitatively assess the image reproduction at far, intermediate, and near distances, offering a visual representation of the IOLs’ capabilities. Conclusions: These findings provide valuable information for selecting IOLs based on patient-specific visual requirements in cataract and refractive surgery.| File | Dimensione | Formato | |
|---|---|---|---|
|
ART xx Journal of Optometry 2025 .pdf
accesso aperto
Tipologia:
Pdf editoriale (Version of record)
Licenza:
Open Access
Dimensione
1.15 MB
Formato
Adobe PDF
|
1.15 MB | Adobe PDF |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.



