This study reports the measurement, using the Lambda-X PMTF and Lambda-X instruments, of the optical characteristics of three types of intraocular lenses (IOLs) and the subsequent analysis for evaluation and comparison. The optical characteristics of the lenses are evaluated through the Modulation Transfer Function (MTF) in Through Focus mode and with different pupil sizes. The USAF (United States Air Force) resolution target is also used to qualitatively assess the image reproduction at far, intermediate, and near distances, offering a visual representation of the IOLs’ capabilities. The results show different behavior of the lenses in the visual acuity at different distances. In addition, the data were used in combination with previously reported proposals to connect the MTF with postoperative visual acuity.
Quantification of IOLs’ characteristics through optical measurements: Indications for a pre-implantation vision assessment at different distances / Alessio Giorgetti, Federico Tommasi, Lorenzo Fini, Stefano Pelli, Rita Mencucci, Giovanni Romualdi, Massimo Gurioli, Stefano Cavalieri. - In: IL NUOVO CIMENTO C. - ISSN 2037-4909. - ELETTRONICO. - 48:(2025), pp. 262.0-262.0. [10.1393/ncc/i2025-25262-3]
Quantification of IOLs’ characteristics through optical measurements: Indications for a pre-implantation vision assessment at different distances
Alessio Giorgetti;Federico Tommasi
;Lorenzo Fini;Stefano Pelli;Rita Mencucci;Giovanni Romualdi;Massimo Gurioli;Stefano Cavalieri
2025
Abstract
This study reports the measurement, using the Lambda-X PMTF and Lambda-X instruments, of the optical characteristics of three types of intraocular lenses (IOLs) and the subsequent analysis for evaluation and comparison. The optical characteristics of the lenses are evaluated through the Modulation Transfer Function (MTF) in Through Focus mode and with different pupil sizes. The USAF (United States Air Force) resolution target is also used to qualitatively assess the image reproduction at far, intermediate, and near distances, offering a visual representation of the IOLs’ capabilities. The results show different behavior of the lenses in the visual acuity at different distances. In addition, the data were used in combination with previously reported proposals to connect the MTF with postoperative visual acuity.| File | Dimensione | Formato | |
|---|---|---|---|
|
ncc13426.pdf
accesso aperto
Tipologia:
Versione finale referata (Postprint, Accepted manuscript)
Licenza:
Creative commons
Dimensione
921.19 kB
Formato
Adobe PDF
|
921.19 kB | Adobe PDF |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.