This study describes the analysis of microplastics in aqueous matrices of varying complexity using pyrolysis coupled with gas chromatography-mass spectrometry (Py-GC–MS). Depending on the filtration rate of the analysed matrix, sample treatment prior to Py-GC–MS consisted of filtration on 0.7 μm glass fibre membranes (low-complexity matrices), or microwave-assisted digestion of samples with hydrogen peroxide, followed by filtration as above-mentioned (high-complexity matrices). Low-complexity matrices included demineralised water (DMW), mineral water in polyethylene terephthalate (PET) bottles (BMW), drinking water from a public aqueduct (DW), and DW further treated with activated carbon and distributed in public fountains (FOW). Highcomplexity matrices included effluent wastewater from primary sedimentation (WW-PS), biological oxidation (WW-BO), clariflocculation (WW-CL), and quaternary treatment (WW-QT). Sample digestion significantly accelerated the filtration process of high-complexity matrices. In most cases, absolute value of matrix effect was <40 %, except for PET in WW-PS (-54 %) and WW-CL (-56 %), and polystyrene in WW-PS (75 %). Apparent recovery, studied in DMW, DW, and WW-BO matrices, was 90±24, 87±25, and 67±19, respectively. The methods were evaluated for their greenness, achieving scores higher than or comparable to those of previously published Py-GC–MS procedures. Poly(methyl methacrylate) and PET were never detected, while different levels of contamination were found for polystyrene, polyethylene, polypropylene, polycarbonate, polyamide, and polyvinyl chloride, depending on the sample type. The lowest contaminations were determined in BMW and DMW. FOW was more contaminated than DW (five microplastics detected at 2.7–138 μg/L vs. two microplastics detected at <0.62–23 μg/L) and concentrations in WWs were in the order WW-QT<
Identification and quantification of microplastics in a wide range of aqueous matrices using pyrolysis-gas chromatography-mass spectrometry: method’s performance, greenness evaluation, and analysis of real samples / Giulia Bonaccorso, Simone Bonari, Paolo Bini, Christine Savorgianaki, Lapo Renai, Michelangelo Fichera, Donatella Fibbi, Daniela Santianni, Claudia Bello, Anna Maria Papini, Massimo Del Bubba. - In: GREEN ANALYTICAL CHEMISTRY. - ISSN 2772-5774. - ELETTRONICO. - 15:(2025), pp. 100301.0-100301.0. [10.1016/j.greeac.2025.100301]
Identification and quantification of microplastics in a wide range of aqueous matrices using pyrolysis-gas chromatography-mass spectrometry: method’s performance, greenness evaluation, and analysis of real samples
Giulia Bonaccorso;Paolo Bini;Lapo Renai;Michelangelo Fichera;Donatella Fibbi;Claudia Bello;Anna Maria Papini;Massimo Del Bubba
2025
Abstract
This study describes the analysis of microplastics in aqueous matrices of varying complexity using pyrolysis coupled with gas chromatography-mass spectrometry (Py-GC–MS). Depending on the filtration rate of the analysed matrix, sample treatment prior to Py-GC–MS consisted of filtration on 0.7 μm glass fibre membranes (low-complexity matrices), or microwave-assisted digestion of samples with hydrogen peroxide, followed by filtration as above-mentioned (high-complexity matrices). Low-complexity matrices included demineralised water (DMW), mineral water in polyethylene terephthalate (PET) bottles (BMW), drinking water from a public aqueduct (DW), and DW further treated with activated carbon and distributed in public fountains (FOW). Highcomplexity matrices included effluent wastewater from primary sedimentation (WW-PS), biological oxidation (WW-BO), clariflocculation (WW-CL), and quaternary treatment (WW-QT). Sample digestion significantly accelerated the filtration process of high-complexity matrices. In most cases, absolute value of matrix effect was <40 %, except for PET in WW-PS (-54 %) and WW-CL (-56 %), and polystyrene in WW-PS (75 %). Apparent recovery, studied in DMW, DW, and WW-BO matrices, was 90±24, 87±25, and 67±19, respectively. The methods were evaluated for their greenness, achieving scores higher than or comparable to those of previously published Py-GC–MS procedures. Poly(methyl methacrylate) and PET were never detected, while different levels of contamination were found for polystyrene, polyethylene, polypropylene, polycarbonate, polyamide, and polyvinyl chloride, depending on the sample type. The lowest contaminations were determined in BMW and DMW. FOW was more contaminated than DW (five microplastics detected at 2.7–138 μg/L vs. two microplastics detected at <0.62–23 μg/L) and concentrations in WWs were in the order WW-QT<| File | Dimensione | Formato | |
|---|---|---|---|
|
Green Analytical Chemistry 2025.pdf
accesso aperto
Tipologia:
Pdf editoriale (Version of record)
Licenza:
Open Access
Dimensione
1.54 MB
Formato
Adobe PDF
|
1.54 MB | Adobe PDF |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.