Background and aims: The use of a POCT (Point Of Care Test) could help in reducing the impact of pre-analytical errors in particular in challenging newborn samples. The study purpose is to compare the POCT Sight OLO® hematology analyzer, validated for >3 months patients, with the reference system Sysmex XN-9100™ in Neonatal Intensive Care Unit (NICU). Material and methods: The two analyzers were compared through Passing-Bablok regression analysis and Bland-Altman plot. Results: We analyzed 65 blood samples, in detail 38 from adults and 27 from newborns. The regression analysis results performed in the newborn and adult patients showed a good agreement between the two instruments. The evaluation of the Bland-Altman plots showed comparable values of bias <10 % for the most of parameters. The evaluation of sample flags for the presence of distributional and morphological abnormalities showed a partial accordance between the two approaches, but the POCT exhibited good performance compared to the final report revised by the laboratory specialist. Conclusions: The comparison of the two instruments demonstrated that they provide comparable blood counts, also in patients aged <3 months. The POCT allows having reliable analytical data and faster turning around time, particularly useful in NICU.
Complete blood count in neonatal Intensive care Unit (NICU): Performance comparison between POCT Sight OLO® and Sysmex XN-9100™ hematology analyzers / Nencini, Francesca; Bonari, Alessandro; Ciullini Mannurita, Sara; Mongia, Alessandra; Romano, Francesca; Garieri, Maria; Russo, Edda; Sastrucci, Silvia; Marrani, Graziella; Tonelli, Martina; Salti, Stefano; Funel, Niccola; Amedei, Amedeo; Dani, Carlo; Fanelli, Alessandra. - In: PRACTICAL LABORATORY MEDICIN. - ISSN 2352-5517. - ELETTRONICO. - 44:(2025), pp. e00453.0-e00453.0. [10.1016/j.plabm.2025.e00453]
Complete blood count in neonatal Intensive care Unit (NICU): Performance comparison between POCT Sight OLO® and Sysmex XN-9100™ hematology analyzers
Nencini, Francesca;Bonari, Alessandro;Ciullini Mannurita, Sara;Garieri, Maria;Russo, Edda;Salti, Stefano;Amedei, Amedeo;Dani, Carlo;Fanelli, Alessandra
2025
Abstract
Background and aims: The use of a POCT (Point Of Care Test) could help in reducing the impact of pre-analytical errors in particular in challenging newborn samples. The study purpose is to compare the POCT Sight OLO® hematology analyzer, validated for >3 months patients, with the reference system Sysmex XN-9100™ in Neonatal Intensive Care Unit (NICU). Material and methods: The two analyzers were compared through Passing-Bablok regression analysis and Bland-Altman plot. Results: We analyzed 65 blood samples, in detail 38 from adults and 27 from newborns. The regression analysis results performed in the newborn and adult patients showed a good agreement between the two instruments. The evaluation of the Bland-Altman plots showed comparable values of bias <10 % for the most of parameters. The evaluation of sample flags for the presence of distributional and morphological abnormalities showed a partial accordance between the two approaches, but the POCT exhibited good performance compared to the final report revised by the laboratory specialist. Conclusions: The comparison of the two instruments demonstrated that they provide comparable blood counts, also in patients aged <3 months. The POCT allows having reliable analytical data and faster turning around time, particularly useful in NICU.| File | Dimensione | Formato | |
|---|---|---|---|
|
1-s2.0-S235255172500006X-main.pdf
accesso aperto
Tipologia:
Pdf editoriale (Version of record)
Licenza:
Open Access
Dimensione
5.09 MB
Formato
Adobe PDF
|
5.09 MB | Adobe PDF |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.