A lateral flow immunoassay (LFIA) device was developed and applied to testing urine samples for residues of the antimicrobial sulphamethazine (SMZ). This report describes the preparation of a rat monoclonal antibody to SMZ and its characterisation in an ELISA format. Apart from SMZ, the antibody showed high ( z 50%) cross-reactivity to N4-acetylsulphamethazine (55%), sulphamerazine (59%) and sulphisoxazole (50%) and lower cross-reactivity of 18% to sulphachlorpyridazine and sulphadiazine. The LFIA device consisted of a nitrocellulose membrane spotted with SMZovalbumin and goat anti-mouse antibody as capture line and control line, respectively. Mouse anti-rat IgG F(abV)2 fragment specific antibody, adsorbed to colloidal carbon, was used as the detection ligand in the LFIA. The LFIA device had a cut-off value of 6.3 ng/ml in diluted (1/10) urine. Urine samples from SMZ-treated pigs, and bovine and porcine urine samples fortified with SMZ were used for a blind, four-laboratory evaluation of the performance of the LFIA device. Concentrations of SMZ in the test samples (n = 29), as determined by LC-MS/MS, ranged from 0 ( < 3) to 1174 ng/ml. The evaluation of the LFIA device showed an overall sensitivity of 100%, a specificity of 71%, and positive and negative prediction values of 73% and 100%, respectively. The LFIA device has been fabricated as a test kit for determining SMZ residues in animals produced for slaughter.

Preliminary evaluation of a lateral flow immunoassay device for screening urine samples for the presence of sulphamethazine / O'KEEFFE M; CRABBE P; SALDEN M; WICHERS J; VAN PETEGHEM C; KOHEN F; PIERACCINI G; G. MONETI. - In: JOURNAL OF IMMUNOLOGICAL METHODS. - ISSN 0022-1759. - STAMPA. - 278:(2003), pp. 117-126.

Preliminary evaluation of a lateral flow immunoassay device for screening urine samples for the presence of sulphamethazine

PIERACCINI, GIUSEPPE;MONETI, GLORIANO
2003

Abstract

A lateral flow immunoassay (LFIA) device was developed and applied to testing urine samples for residues of the antimicrobial sulphamethazine (SMZ). This report describes the preparation of a rat monoclonal antibody to SMZ and its characterisation in an ELISA format. Apart from SMZ, the antibody showed high ( z 50%) cross-reactivity to N4-acetylsulphamethazine (55%), sulphamerazine (59%) and sulphisoxazole (50%) and lower cross-reactivity of 18% to sulphachlorpyridazine and sulphadiazine. The LFIA device consisted of a nitrocellulose membrane spotted with SMZovalbumin and goat anti-mouse antibody as capture line and control line, respectively. Mouse anti-rat IgG F(abV)2 fragment specific antibody, adsorbed to colloidal carbon, was used as the detection ligand in the LFIA. The LFIA device had a cut-off value of 6.3 ng/ml in diluted (1/10) urine. Urine samples from SMZ-treated pigs, and bovine and porcine urine samples fortified with SMZ were used for a blind, four-laboratory evaluation of the performance of the LFIA device. Concentrations of SMZ in the test samples (n = 29), as determined by LC-MS/MS, ranged from 0 ( < 3) to 1174 ng/ml. The evaluation of the LFIA device showed an overall sensitivity of 100%, a specificity of 71%, and positive and negative prediction values of 73% and 100%, respectively. The LFIA device has been fabricated as a test kit for determining SMZ residues in animals produced for slaughter.
2003
278
117
126
O'KEEFFE M; CRABBE P; SALDEN M; WICHERS J; VAN PETEGHEM C; KOHEN F; PIERACCINI G; G. MONETI
File in questo prodotto:
File Dimensione Formato  
2003_ OKeeffe.pdf

Accesso chiuso

Tipologia: Versione finale referata (Postprint, Accepted manuscript)
Licenza: Tutti i diritti riservati
Dimensione 380.67 kB
Formato Adobe PDF
380.67 kB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/215219
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact