The validation of a HPTLC-densitometric method for the determination of secoisolariciresinol diglucoside (SDG) in flaxseed was performed improving the reproducibility of a previously reported HPTLC densitometric procedure by the use of fully wettable reversed phase plates (silica gel 60 RP18W F254S, 10 cm×10 cm) with MeOH:HCOOH 0.1% (40:60, v/v) mobile phase. The analysis required only the alkaline hydrolysis in aqueous medium of undefatted samples and densitometry at 282nm of HPTLC runs. The method was validated following the protocol proposed by the Société Francaise des Sciences et Techniques Pharmaceutiques (SFSTP) giving rise to a dependable and high throughput procedure well suited to routine application. SDG was quantified in the range of 321–1071 ng with RSD of repeatability and intermediate precision not exceeding 3.61% and accuracy inside the acceptance limits. Flaxseed of five cultivars of different origin was elected as test-bed.
Validation of a reversed phase high performance thin layer chromatographic-densitometric method for secoisolariciresinol diglucoside determination in flaxseed / Silvia A Coran; Gianluca Bartolucci; Massimo Bambagiotti-Alberti. - In: JOURNAL OF CHROMATOGRAPHY A. - ISSN 0021-9673. - STAMPA. - 1207:(2008), pp. 155-159. [10.1016/j.chroma.2008.08.074]
Validation of a reversed phase high performance thin layer chromatographic-densitometric method for secoisolariciresinol diglucoside determination in flaxseed.
CORAN, SILVIA;BARTOLUCCI, GIAN LUCA;BAMBAGIOTTI ALBERTI, MASSIMO
2008
Abstract
The validation of a HPTLC-densitometric method for the determination of secoisolariciresinol diglucoside (SDG) in flaxseed was performed improving the reproducibility of a previously reported HPTLC densitometric procedure by the use of fully wettable reversed phase plates (silica gel 60 RP18W F254S, 10 cm×10 cm) with MeOH:HCOOH 0.1% (40:60, v/v) mobile phase. The analysis required only the alkaline hydrolysis in aqueous medium of undefatted samples and densitometry at 282nm of HPTLC runs. The method was validated following the protocol proposed by the Société Francaise des Sciences et Techniques Pharmaceutiques (SFSTP) giving rise to a dependable and high throughput procedure well suited to routine application. SDG was quantified in the range of 321–1071 ng with RSD of repeatability and intermediate precision not exceeding 3.61% and accuracy inside the acceptance limits. Flaxseed of five cultivars of different origin was elected as test-bed.File | Dimensione | Formato | |
---|---|---|---|
SDG fulltext.pdf
Accesso chiuso
Tipologia:
Altro
Licenza:
Tutti i diritti riservati
Dimensione
526.31 kB
Formato
Adobe PDF
|
526.31 kB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.