BACKGROUND: To define the best strategy to reduce Compound A production in Sevoflurane low-flow anaesthesia by experiments in vitro and in vivo of different absorbers and different anaesthesia machines. METHODS: In vitro Compound A has been measured at 45 degrees C in vitro following Sevoflurane interactions with potassium hydroxide, sodium hydroxide, soda lime, Dragersorb 800 Plus and Amsorb, a new absorber that does not contain sodium or potassium hydroxide. In vivo Compound A concentration in the anaesthesia circuit (inspiratory branch) has been measured using an indirect sampling method through absorber vials (SKC) with active coal granules, during low flows (500 ml/min) general anaesthesia using soda lime, Dragersorb 800 Plus or Amsorb as absorber. Compound A was also measured during low flows (500 ml/min) general anaesthesia using as carbon dioxide absorber soda lime with different anaesthesia machines. RESULTS: In vitro at 45 degrees C Compound A concentration with soda lime and Dragersorb 800 Plus was about 10 times higher than with Amsorb. In vivo the Compound A concentrations in the inspiratory branch of the circuit were lower in the group with Amsorb. CONCLUSION: The Compound A production is minimal with Amsorb as carbon dioxide absorber.

SEVOFLURANE LOW-FLOW ANAESTHESIA: BEST STRATEGY TO REDUCE COMPOUND A CONCENTRATION / A.Di Filippo;F.Marini;M.PACENTI;S.DUGHERI;L.FOCARDI;G.P.NOVELLI. - In: ACTA ANAESTHESIOLOGICA SCANDINAVICA. - ISSN 1399-6576. - STAMPA. - 46:(2002), pp. 1017-1020.

SEVOFLURANE LOW-FLOW ANAESTHESIA: BEST STRATEGY TO REDUCE COMPOUND A CONCENTRATION

DI FILIPPO, ALESSANDRO;NOVELLI, GIAN PAOLO
2002

Abstract

BACKGROUND: To define the best strategy to reduce Compound A production in Sevoflurane low-flow anaesthesia by experiments in vitro and in vivo of different absorbers and different anaesthesia machines. METHODS: In vitro Compound A has been measured at 45 degrees C in vitro following Sevoflurane interactions with potassium hydroxide, sodium hydroxide, soda lime, Dragersorb 800 Plus and Amsorb, a new absorber that does not contain sodium or potassium hydroxide. In vivo Compound A concentration in the anaesthesia circuit (inspiratory branch) has been measured using an indirect sampling method through absorber vials (SKC) with active coal granules, during low flows (500 ml/min) general anaesthesia using soda lime, Dragersorb 800 Plus or Amsorb as absorber. Compound A was also measured during low flows (500 ml/min) general anaesthesia using as carbon dioxide absorber soda lime with different anaesthesia machines. RESULTS: In vitro at 45 degrees C Compound A concentration with soda lime and Dragersorb 800 Plus was about 10 times higher than with Amsorb. In vivo the Compound A concentrations in the inspiratory branch of the circuit were lower in the group with Amsorb. CONCLUSION: The Compound A production is minimal with Amsorb as carbon dioxide absorber.
2002
46
1017
1020
A.Di Filippo;F.Marini;M.PACENTI;S.DUGHERI;L.FOCARDI;G.P.NOVELLI
File in questo prodotto:
File Dimensione Formato  
di filippo sevo 2002.pdf

accesso aperto

Tipologia: Versione finale referata (Postprint, Accepted manuscript)
Licenza: Open Access
Dimensione 49.11 kB
Formato Adobe PDF
49.11 kB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/330715
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 7
  • ???jsp.display-item.citation.isi??? 5
social impact