The possibility of direct liquid injection inside an ion trap, designed for a commercial gas chromatography/mass spectrometry (GC/MS) system, has been tested by continuous solution flow and by loop injection. Instrumental parameters such as solution flows and sample concentration, and temperatures of transfer line, ion trap and ion trap housing, have been optimised. Different configurations of the interface capillary have been also considered, and the best results have been obtained by the injection of the solution coaxially with a helium flow. The experimental arrangement leads to the detection of good quality electron (EI) and chemical (CI) ionisation spectra of low to medium polarity compounds, at µM levels. The direct coupling of a nano-HPLC column with the modified transfer line led on one hand to EI spectra which were successfully employed for library searches, and on the other to CI data directly comparable to results achieved by a commercial HPLC/MS system operating under ESI conditions.

Direct coupling of a nano-high-performance liquid chromatography column to an ion trap designed for a gas chromatography/mass spectrometry system / G. Moneti; G. Pieraccini; G. Bartolucci; G. Guida; R. Seraglia; P. Traldi. - In: RAPID COMMUNICATIONS IN MASS SPECTROMETRY. - ISSN 0951-4198. - ELETTRONICO. - 15:(2001), pp. 1607-1617. [10.1002/rcm.407]

Direct coupling of a nano-high-performance liquid chromatography column to an ion trap designed for a gas chromatography/mass spectrometry system

MONETI, GLORIANO;PIERACCINI, GIUSEPPE;BARTOLUCCI, GIAN LUCA;
2001

Abstract

The possibility of direct liquid injection inside an ion trap, designed for a commercial gas chromatography/mass spectrometry (GC/MS) system, has been tested by continuous solution flow and by loop injection. Instrumental parameters such as solution flows and sample concentration, and temperatures of transfer line, ion trap and ion trap housing, have been optimised. Different configurations of the interface capillary have been also considered, and the best results have been obtained by the injection of the solution coaxially with a helium flow. The experimental arrangement leads to the detection of good quality electron (EI) and chemical (CI) ionisation spectra of low to medium polarity compounds, at µM levels. The direct coupling of a nano-HPLC column with the modified transfer line led on one hand to EI spectra which were successfully employed for library searches, and on the other to CI data directly comparable to results achieved by a commercial HPLC/MS system operating under ESI conditions.
2001
15
1607
1617
G. Moneti; G. Pieraccini; G. Bartolucci; G. Guida; R. Seraglia; P. Traldi
File in questo prodotto:
File Dimensione Formato  
RCM2001.pdf

Accesso chiuso

Tipologia: Versione finale referata (Postprint, Accepted manuscript)
Licenza: Tutti i diritti riservati
Dimensione 306.3 kB
Formato Adobe PDF
306.3 kB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/335475
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 4
social impact