Purpose Conflicting data exist about the difference between 8- and 16-frame gated single-photon emission computed tomography (SPECT) left ventricular volumes and ejection fraction (EF); moreover, the influence of framing on detection of stress-induced functional changes is unknown. Methods In 133 patients, two separate gated SPECT studies, one with 8 and one with 16 frames, were simultaneously acquired during a single gantry orbit using dedicated software. In 33 of 133 patients, two additional studies (with 8 and 16 frames, respectively) were acquired using arrhythmia rejection. Left ventricular EF and volumes were calculated using the QGS software. Stress-induced ischemia was identified on summed perfusion images. Results Arrhythmia-rejection did not influence volumes and EF independently of framing rate. Using data without arrhythmia-rejection, there was a significant difference in volumes and EF between 8 and 16 frames both in resting and post-stress gated SPECT. However, the difference was small: 2.6% for resting and 2.8% for post-stress EF. Both using 8 and 16 frames, there were significantly larger volumes and lower EF in patients with than without stress-induced ischemia. A stress-induced decrease >5 EF units was observed in 26 of 133 patients using 8 and in 23 of 133 using 16 frames, respectively, with finding agreement in 19 patients. Conclusions Comparing two simultaneously acquired studies, the use of 16 instead of 8 frames has minor and predictable influence on functional data. Furthermore, there are no differences in the detection of stress-induced functional changes. The advantage of 16 over 8 frames in the daily clinical practice appears questionable.
Is 16-frame really superior to 8-frame gated SPECT for the assessment of left ventricular volumes and ejection fraction? Comparison of two simultaneously acquired gated SPECT studies / Giulia Montelatici; Roberto Sciagrà; Alessandro Passeri; Manjola Dona; Alberto Pupi. - In: EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING. - ISSN 1619-7070. - STAMPA. - 35:(2008), pp. 2059-2065. [10.1007/s00259-008-0866-2]
Is 16-frame really superior to 8-frame gated SPECT for the assessment of left ventricular volumes and ejection fraction? Comparison of two simultaneously acquired gated SPECT studies.
SCIAGRA', ROBERTO;PASSERI, ALESSANDRO;PUPI, ALBERTO
2008
Abstract
Purpose Conflicting data exist about the difference between 8- and 16-frame gated single-photon emission computed tomography (SPECT) left ventricular volumes and ejection fraction (EF); moreover, the influence of framing on detection of stress-induced functional changes is unknown. Methods In 133 patients, two separate gated SPECT studies, one with 8 and one with 16 frames, were simultaneously acquired during a single gantry orbit using dedicated software. In 33 of 133 patients, two additional studies (with 8 and 16 frames, respectively) were acquired using arrhythmia rejection. Left ventricular EF and volumes were calculated using the QGS software. Stress-induced ischemia was identified on summed perfusion images. Results Arrhythmia-rejection did not influence volumes and EF independently of framing rate. Using data without arrhythmia-rejection, there was a significant difference in volumes and EF between 8 and 16 frames both in resting and post-stress gated SPECT. However, the difference was small: 2.6% for resting and 2.8% for post-stress EF. Both using 8 and 16 frames, there were significantly larger volumes and lower EF in patients with than without stress-induced ischemia. A stress-induced decrease >5 EF units was observed in 26 of 133 patients using 8 and in 23 of 133 using 16 frames, respectively, with finding agreement in 19 patients. Conclusions Comparing two simultaneously acquired studies, the use of 16 instead of 8 frames has minor and predictable influence on functional data. Furthermore, there are no differences in the detection of stress-induced functional changes. The advantage of 16 over 8 frames in the daily clinical practice appears questionable.File | Dimensione | Formato | |
---|---|---|---|
articolo_1.pdf
Accesso chiuso
Tipologia:
Altro
Licenza:
Tutti i diritti riservati
Dimensione
220.23 kB
Formato
Adobe PDF
|
220.23 kB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.