The PIXE technique is a reliable tool for the characterisation of thin aerosol samples, but it can underestimate the lightest measurable elements, like Na, Mg, Al, Si and P, owing to the absorption of their X-rays inside the sample. The PIGE technique is a valid help to determine corrections for such effect: in order to perform PIGE measurements relative to thin reference standards in an external beam set-up, we measured, at the external beam facility of the Tandetron accelerator of the LABEC laboratory in Florence, the g-ray yields as a function of the proton beam energy for the reactions 19F(p, p'g)19F (Eg = 110 and 197 keV), 23Na(p,p'g)23Na (Eg = 440 keV) and 27Al(p, p'g)27Al (Eg = 843 and 1013 keV), in the proton energy range from 3 to 5 MeV. The measured yields are shown, and the determined most suitable energies for performing PIGE quantification of Na and Al are reported, together with the corresponding minimum detection limits (MDLs). The results of some test on PIGE accuracy and an evaluation of self-absorption effects in PIXE measurements on thin aerosol samples are also presented.
Proton induced gamma-ray emission yields for the analysis of light elements in aerosol samples in an external beam set-up / G. Calzolai; M. Chiari; F. Lucarelli; S. Nava; S. Portarena. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION B, BEAM INTERACTIONS WITH MATERIALS AND ATOMS. - ISSN 0168-583X. - STAMPA. - 268:(2010), pp. 1540-1545. [10.1016/j.nimb.2010.03.002]
Proton induced gamma-ray emission yields for the analysis of light elements in aerosol samples in an external beam set-up
CALZOLAI, GIULIA;CHIARI, MASSIMO;LUCARELLI, FRANCO;NAVA, SILVIA;
2010
Abstract
The PIXE technique is a reliable tool for the characterisation of thin aerosol samples, but it can underestimate the lightest measurable elements, like Na, Mg, Al, Si and P, owing to the absorption of their X-rays inside the sample. The PIGE technique is a valid help to determine corrections for such effect: in order to perform PIGE measurements relative to thin reference standards in an external beam set-up, we measured, at the external beam facility of the Tandetron accelerator of the LABEC laboratory in Florence, the g-ray yields as a function of the proton beam energy for the reactions 19F(p, p'g)19F (Eg = 110 and 197 keV), 23Na(p,p'g)23Na (Eg = 440 keV) and 27Al(p, p'g)27Al (Eg = 843 and 1013 keV), in the proton energy range from 3 to 5 MeV. The measured yields are shown, and the determined most suitable energies for performing PIGE quantification of Na and Al are reported, together with the corresponding minimum detection limits (MDLs). The results of some test on PIGE accuracy and an evaluation of self-absorption effects in PIXE measurements on thin aerosol samples are also presented.File | Dimensione | Formato | |
---|---|---|---|
Proton induced gamma-ray emission yields for the analysis of light elements in aerosol samples in an external beam set-up.pdf
Accesso chiuso
Tipologia:
Versione finale referata (Postprint, Accepted manuscript)
Licenza:
Tutti i diritti riservati
Dimensione
597.81 kB
Formato
Adobe PDF
|
597.81 kB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.