The need for investigating the properties of new materials at nanoscale level continuously pushes the development of higher resolution measurement instruments. In this context, a promising dynamic atomic force microscopy setup, where the cantilever gets excited by a nonlinear feedback loop, has been recently introduced. In the first part of the paper, the application of this working mode to imaging is experimentally investigated, showing the effectiveness of this novel approach. Furthermore, the presence of a variable saturation in the nonlinear loop is exploited to design a specific algorithm that dynamically adapts the cantilever free oscillation amplitude to sudden variations of the sample profile. In imaging applications, this additional control action significantly reduces the tip–sample interaction force yet maintaining high image quality, thus resulting in a suitable setup for better preserving the state of soft and damageable samples such as biological specimens.
AFM Imaging via Nonlinear Control of Self-driven Cantilever Oscillations / M. Basso;P. Paoletti;B. Tiribilli;M. Vassalli. - In: IEEE TRANSACTIONS ON NANOTECHNOLOGY. - ISSN 1536-125X. - STAMPA. - 10:(2011), pp. 560-565. [10.1109/TNANO.2010.2051815]
AFM Imaging via Nonlinear Control of Self-driven Cantilever Oscillations
BASSO, MICHELE;PAOLETTI, PAOLO;VASSALLI, MASSIMO
2011
Abstract
The need for investigating the properties of new materials at nanoscale level continuously pushes the development of higher resolution measurement instruments. In this context, a promising dynamic atomic force microscopy setup, where the cantilever gets excited by a nonlinear feedback loop, has been recently introduced. In the first part of the paper, the application of this working mode to imaging is experimentally investigated, showing the effectiveness of this novel approach. Furthermore, the presence of a variable saturation in the nonlinear loop is exploited to design a specific algorithm that dynamically adapts the cantilever free oscillation amplitude to sudden variations of the sample profile. In imaging applications, this additional control action significantly reduces the tip–sample interaction force yet maintaining high image quality, thus resulting in a suitable setup for better preserving the state of soft and damageable samples such as biological specimens.File | Dimensione | Formato | |
---|---|---|---|
tnano10.pdf
Accesso chiuso
Tipologia:
Versione finale referata (Postprint, Accepted manuscript)
Licenza:
Tutti i diritti riservati
Dimensione
596.2 kB
Formato
Adobe PDF
|
596.2 kB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.