Test the validity of jitter measurement in strongly irregular voice signals. • Four tools are compared on synthetic noisy signals corrupted by increasing jitter. • One method (BioVoice) allows estimating jitter values up to and larger than 20%. • BioVoice can give a valid support to clinicians.
Perturbation measurements in highly irregular voice signals: Performances/validity of analysis software tools / C. Manfredi; A. Giordano; J. Schoentgen; S. Fraj; L. Bocchi; P.H. Dejonckere. - In: BIOMEDICAL SIGNAL PROCESSING AND CONTROL. - ISSN 1746-8094. - STAMPA. - 7:(2012), pp. 409-416. [10.1016/j.bspc.2011.06.004]
Perturbation measurements in highly irregular voice signals: Performances/validity of analysis software tools
MANFREDI, CLAUDIA;BOCCHI, LEONARDO;
2012
Abstract
Test the validity of jitter measurement in strongly irregular voice signals. • Four tools are compared on synthetic noisy signals corrupted by increasing jitter. • One method (BioVoice) allows estimating jitter values up to and larger than 20%. • BioVoice can give a valid support to clinicians.File in questo prodotto: