commonly used penile prostheses (the American Medical Systems [AMS] series). AIM: The study aimed to assess, retrospectively, complications and patient satisfaction with AMS penile implants in 253 consecutive patients with erectile dysfunction from three European centers. METHODS: Pre, intra- and postoperative data were obtained from chart review, with a mean follow-up of 60 months; 200 patients were available for evaluation. Patient satisfaction data were collected using the Erectile Dysfunction Inventory of Treatment Satisfaction (EDITS) questionnaire. MAIN OUTCOME MEASURE: Complications and patient satisfaction were assessed. Patient satisfaction was evaluated using a standardized assessment tool (the modified EDITS questionnaire). RESULTS: Major postoperative complications occurred in 40 (20%) patients, including 9 (22.5%) prosthesis infections, 18 (45%) mechanical failures, and 13 (32.5%) erosions. Patient satisfaction with the AMS 700CX, AMS Ambicor, and AMS 600-650 was 97%, 81%, and 75%, respectively; dissatisfaction was 0%, 5%, and 6%, respectively. Partner satisfaction with the AMS 700CX, AMS Ambicor, and AMS 600-650 was 91%, 91%, and 75%, respectively; dissatisfaction was 0%, 5%, and 6%, respectively. Erections were more natural (harder) than before with the AMS 700CX, AMS Ambicor, and AMS 600-650 in 91%, 85%, and 88%, respectively; hardness was the same as before in 9%, 15%, and 13%, respectively; no erections were less hard than before. CONCLUSIONS: Postoperative complications differed from those reported in the literature, while patient satisfaction rates were roughly similar. The reporting of specific data for different implant types, plus the use of standardized assessment tools for patient satisfaction is significant as in the future, it will allow comparison of data between centers performing penile prosthesis implants using these devices.
Penile implantation in Europe: successes and complications with 253 implants in Italy and Germany / Natali A;Olianas R;Fisch M. - In: JOURNAL OF SEXUAL MEDICINE. - ISSN 1743-6095. - ELETTRONICO. - 5:(2008), pp. 1503-1512. [10.1111/j.1743-6109.2008.00819.x]
Penile implantation in Europe: successes and complications with 253 implants in Italy and Germany
NATALI, ALESSANDRO;
2008
Abstract
commonly used penile prostheses (the American Medical Systems [AMS] series). AIM: The study aimed to assess, retrospectively, complications and patient satisfaction with AMS penile implants in 253 consecutive patients with erectile dysfunction from three European centers. METHODS: Pre, intra- and postoperative data were obtained from chart review, with a mean follow-up of 60 months; 200 patients were available for evaluation. Patient satisfaction data were collected using the Erectile Dysfunction Inventory of Treatment Satisfaction (EDITS) questionnaire. MAIN OUTCOME MEASURE: Complications and patient satisfaction were assessed. Patient satisfaction was evaluated using a standardized assessment tool (the modified EDITS questionnaire). RESULTS: Major postoperative complications occurred in 40 (20%) patients, including 9 (22.5%) prosthesis infections, 18 (45%) mechanical failures, and 13 (32.5%) erosions. Patient satisfaction with the AMS 700CX, AMS Ambicor, and AMS 600-650 was 97%, 81%, and 75%, respectively; dissatisfaction was 0%, 5%, and 6%, respectively. Partner satisfaction with the AMS 700CX, AMS Ambicor, and AMS 600-650 was 91%, 91%, and 75%, respectively; dissatisfaction was 0%, 5%, and 6%, respectively. Erections were more natural (harder) than before with the AMS 700CX, AMS Ambicor, and AMS 600-650 in 91%, 85%, and 88%, respectively; hardness was the same as before in 9%, 15%, and 13%, respectively; no erections were less hard than before. CONCLUSIONS: Postoperative complications differed from those reported in the literature, while patient satisfaction rates were roughly similar. The reporting of specific data for different implant types, plus the use of standardized assessment tools for patient satisfaction is significant as in the future, it will allow comparison of data between centers performing penile prosthesis implants using these devices.File | Dimensione | Formato | |
---|---|---|---|
Penile Implants PDF reprints.pdf
Accesso chiuso
Tipologia:
Versione finale referata (Postprint, Accepted manuscript)
Licenza:
Tutti i diritti riservati
Dimensione
113.68 kB
Formato
Adobe PDF
|
113.68 kB | Adobe PDF | Richiedi una copia |
I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.