In this contribution we up-to-date the status of the PRESS-MAG-O device, a new instrument under commissioning at the INFN designed to perform magnetic and spectroscopic experiments on samples under extreme conditions. The system has been designed to work at SINBAD, the IR synchrotron radiation beamline operational at DAFNE. The instrument, that is the result of a significant R&D, will allow performing concurrent high harmonic ac magnetic susceptibility measurements and magneto optic experiments on a sample under high pressure, with a variable DC magnetic field in a wide temperature range. The vacuum vessel has been designed with four crossing windows to allow optical measurements in the transmission geometry on the sample loaded inside a Diamond Anvil Cell. A new superconducting miniaturized micro-SQUID gradiometer has been also developed to detect the low magnetic signal of the sample and a customized optical system has also been designed to perform IR synchrotron radiation experiments.

Status of PRESS-MAG-O: The experimental apparatus to probe materials and phenomena under extreme conditions at Frascati / Di Gioacchino, D.; Marcelli, A.; Cestelli Guidi, M.; Piccinini, M.; Puri, A.; Postorino, P.; Pace, Emanuele; DE SIO, Antonio; Gambicorti, L.. - In: JOURNAL OF PHYSICS AND CHEMISTRY OF SOLIDS. - ISSN 0022-3697. - STAMPA. - 71:(2010), pp. 1042-1045.

Status of PRESS-MAG-O: The experimental apparatus to probe materials and phenomena under extreme conditions at Frascati

PACE, EMANUELE;DE SIO, ANTONIO;
2010

Abstract

In this contribution we up-to-date the status of the PRESS-MAG-O device, a new instrument under commissioning at the INFN designed to perform magnetic and spectroscopic experiments on samples under extreme conditions. The system has been designed to work at SINBAD, the IR synchrotron radiation beamline operational at DAFNE. The instrument, that is the result of a significant R&D, will allow performing concurrent high harmonic ac magnetic susceptibility measurements and magneto optic experiments on a sample under high pressure, with a variable DC magnetic field in a wide temperature range. The vacuum vessel has been designed with four crossing windows to allow optical measurements in the transmission geometry on the sample loaded inside a Diamond Anvil Cell. A new superconducting miniaturized micro-SQUID gradiometer has been also developed to detect the low magnetic signal of the sample and a customized optical system has also been designed to perform IR synchrotron radiation experiments.
2010
71
1042
1045
Di Gioacchino, D.; Marcelli, A.; Cestelli Guidi, M.; Piccinini, M.; Puri, A.; Postorino, P.; Pace, Emanuele; DE SIO, Antonio; Gambicorti, L.
File in questo prodotto:
File Dimensione Formato  
2010 Jour Phys Chem Solids Presmago.pdf

Accesso chiuso

Tipologia: Versione finale referata (Postprint, Accepted manuscript)
Licenza: Tutti i diritti riservati
Dimensione 1.06 MB
Formato Adobe PDF
1.06 MB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/627642
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact