In this work the performance of a prototype three-dimensional (3D) PW Doppler system producing 3D velocity field maps is analyzed by means of measurements on phantoms with known velocity profiles and of simulations. The developed 3-D Doppler system gives a measurement of the velocity vector in a sample volume by means of a probe consisting of four confocal annular array transducers oriented along different directions and three programmable analogue Doppler channels.
Analysis of factors influencing the accuracy of 3D Doppler technique: simulations and experimental results / Calzolai, Marco; Capineri, Lorenzo; Fort, A.; Masotti, Leonardo; Pasqualis, A.; Rocchi, S.; Scabia, Marco. - ELETTRONICO. - (1996), pp. 1245-1248. (Intervento presentato al convegno 1996 IEEE Ultrasonics Symposium tenutosi a San Antonio, Texas) [10.1109/ULTSYM.1996.584219].
Analysis of factors influencing the accuracy of 3D Doppler technique: simulations and experimental results
CALZOLAI, MARCOValidation
;CAPINERI, LORENZO
Conceptualization
;MASOTTI, LEONARDO;SCABIA, MARCO
1996
Abstract
In this work the performance of a prototype three-dimensional (3D) PW Doppler system producing 3D velocity field maps is analyzed by means of measurements on phantoms with known velocity profiles and of simulations. The developed 3-D Doppler system gives a measurement of the velocity vector in a sample volume by means of a probe consisting of four confocal annular array transducers oriented along different directions and three programmable analogue Doppler channels.I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.