Abstract This paper presents a complete PSpice model of an ultrasound single-element transducer, including electrical and mechanical matching as well as the focusing lens. By using this model, it is possible to obtain a relation between the electrical driving source and the acoustic velocity on the transducer surface. This boundary condition then allows the acoustic field to be calculated by numerical methods. Experimental data obtained with two different transducers are in good agreement with results predicted by the related models.
Pspice modelling of ultrasound transducers: comparison of software models to experiment / P. Tortoli; E.Maione; G.Lypacewicz; A.Nowicki; J.Reid. - In: IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL. - ISSN 0885-3010. - STAMPA. - 46:(1999), pp. 399-406. [10.1109/58.753029]
Pspice modelling of ultrasound transducers: comparison of software models to experiment
TORTOLI, PIERO;
1999
Abstract
Abstract This paper presents a complete PSpice model of an ultrasound single-element transducer, including electrical and mechanical matching as well as the focusing lens. By using this model, it is possible to obtain a relation between the electrical driving source and the acoustic velocity on the transducer surface. This boundary condition then allows the acoustic field to be calculated by numerical methods. Experimental data obtained with two different transducers are in good agreement with results predicted by the related models.I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.