Abstract—Mechanical properties of human large arteries result from the interaction between blood pressure, wall distensibility and shear stress. Both the arterial diameter changes through the cardiac cycle (distension) and blood flow velocities can be noninvasively investigated through Doppler ultrasound approaches. Recently, an integrated system processing in real-time all the echo signals produced along an M-line has been developed. This system has been so far demonstrated to be suitable for accurate hemodynamic studies through the detection of blood velocity profiles. This paper reports on the extension of its processing capabilities to the real-time measurement of arterial distension. Tissue motion estimation is based on a modified 2-D autocorrelation algorithm. A novel adaptive approach to track wall position over time using the sum of the high-pass filtered displacement waveform and the low-pass filtered wall position is described. By observing the blood velocity profile, a rapid and accurate positioning of the ultrasound probe and an inherent check on perpendicular observation are provided. First clinical results obtained by measuring the distension of common carotid arteries in a group of 41 volunteers are reported and measurements are validated against those provided by a dedicated wall-track reference system. Average measured distension and diameter were 499 188 m and 6.90 0.66 mm and intraobserver intrasession reproducibility tests showed coefficients of variability of 8.5% and 5.9%, respectively. The agreement between the proposed system and the reference system, expressed as bias 2 SD of the differences, was 34 141 m for distension and 0.05 1.07 mm for diameter.

Clinical validation of common carotid artery wall distension assessment based on multigate Doppler processing / T. Morganti; S. Ricci; F. Vittone; C. Palombo; P. Tortoli. - In: ULTRASOUND IN MEDICINE AND BIOLOGY. - ISSN 0301-5629. - STAMPA. - 31:(2005), pp. 937-945. [10.1016/j.ultrasmedbio.2005.04.001]

Clinical validation of common carotid artery wall distension assessment based on multigate Doppler processing

MORGANTI, TIZIANO;RICCI, STEFANO;TORTOLI, PIERO
2005

Abstract

Abstract—Mechanical properties of human large arteries result from the interaction between blood pressure, wall distensibility and shear stress. Both the arterial diameter changes through the cardiac cycle (distension) and blood flow velocities can be noninvasively investigated through Doppler ultrasound approaches. Recently, an integrated system processing in real-time all the echo signals produced along an M-line has been developed. This system has been so far demonstrated to be suitable for accurate hemodynamic studies through the detection of blood velocity profiles. This paper reports on the extension of its processing capabilities to the real-time measurement of arterial distension. Tissue motion estimation is based on a modified 2-D autocorrelation algorithm. A novel adaptive approach to track wall position over time using the sum of the high-pass filtered displacement waveform and the low-pass filtered wall position is described. By observing the blood velocity profile, a rapid and accurate positioning of the ultrasound probe and an inherent check on perpendicular observation are provided. First clinical results obtained by measuring the distension of common carotid arteries in a group of 41 volunteers are reported and measurements are validated against those provided by a dedicated wall-track reference system. Average measured distension and diameter were 499 188 m and 6.90 0.66 mm and intraobserver intrasession reproducibility tests showed coefficients of variability of 8.5% and 5.9%, respectively. The agreement between the proposed system and the reference system, expressed as bias 2 SD of the differences, was 34 141 m for distension and 0.05 1.07 mm for diameter.
2005
31
937
945
T. Morganti; S. Ricci; F. Vittone; C. Palombo; P. Tortoli
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/312734
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 24
  • ???jsp.display-item.citation.isi??? 26
social impact