Vector Doppler (VD) methods are increasingly attractive for the capability of providing angle-independent blood velocity measurements. This may be particularly important in carotid artery peak systolic velocity (PSV) measurements, which are widely used for stenosis assessments. Here, two recently introduced VD techniques have been compared to the classic spectral Doppler approach through 37 PSV measurements in common carotid arteries (CCAs) and internal carotid arteries (ICAs) of healthy volunteers and of patients with different degrees of stenosis. The PSVs measured by the two VD techniques were very close to each other for both CCAs and ICAs (0.83 slope coefficient and 7 cm/s intercept value, according to the linear regression analysis). Comparing any of the two VD techniques to the classic method provided a (0.42-0.46) regression coefficient with a (14-20) cm/s intercept value. The high correlation obtained with the two VD techniques supports their suitability to perform accurate PSV measurements in carotid arteries

Blood velocity measurement in healthy and diseased carotid arteries by vector Doppler techniques / Matteo Lenge; Daniele Righi; Stefano Ricci; Herve' Liebgott; Piero Tortoli. - ELETTRONICO. - (2014), pp. 345-348. (Intervento presentato al convegno IEEE Ultrasonics Symposium 2014 tenutosi a Chicago nel September 2014) [10.1109/ULTSYM.2014.0085].

Blood velocity measurement in healthy and diseased carotid arteries by vector Doppler techniques

LENGE, MATTEO;RICCI, STEFANO;TORTOLI, PIERO
2014

Abstract

Vector Doppler (VD) methods are increasingly attractive for the capability of providing angle-independent blood velocity measurements. This may be particularly important in carotid artery peak systolic velocity (PSV) measurements, which are widely used for stenosis assessments. Here, two recently introduced VD techniques have been compared to the classic spectral Doppler approach through 37 PSV measurements in common carotid arteries (CCAs) and internal carotid arteries (ICAs) of healthy volunteers and of patients with different degrees of stenosis. The PSVs measured by the two VD techniques were very close to each other for both CCAs and ICAs (0.83 slope coefficient and 7 cm/s intercept value, according to the linear regression analysis). Comparing any of the two VD techniques to the classic method provided a (0.42-0.46) regression coefficient with a (14-20) cm/s intercept value. The high correlation obtained with the two VD techniques supports their suitability to perform accurate PSV measurements in carotid arteries
2014
Proceeding of IEEE Ultrasonics Symposium 2014
IEEE Ultrasonics Symposium 2014
Chicago
September 2014
Matteo Lenge; Daniele Righi; Stefano Ricci; Herve' Liebgott; Piero Tortoli
File in questo prodotto:
File Dimensione Formato  
Proc.pdf

Accesso chiuso

Tipologia: Versione finale referata (Postprint, Accepted manuscript)
Licenza: Tutti i diritti riservati
Dimensione 1.58 MB
Formato Adobe PDF
1.58 MB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/911334
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 1
social impact