Background: Modular dual-mobility (MDM) total hip arthroplasty (THA) is designed with a cobalt-chromium liner inserted into a titanium acetabular component. The purpose of this study is to investigate the potential risks for fretting corrosion at this junction, by measuring serum metal ions, after MDM acetabular revision. Methods: Thirty-seven patients with well-functioning revision THAs participated in a cross-sectional study at mean 5.1 (2-10) years after surgery. All received a trabecular titanium MDM acetabular component. The serum levels of cobalt and chromium were measured using mass spectrometry. Results: The mean values of chromium and cobalt were 2.08 μg/L (95% confidence interval 0.9-3.2, range 0.02-11.8) and 1.99 μg/L (95% confidence interval 0.81-3.17, range 0.07-16.05), respectively.Eleven patients (29, 7%) had ion levels above the normal range, with 6 (16.2%) above 5 μg/L and 5 (13.5%) between 1 and 5 μg/L. A significant correlation was found between an elevated serum metal ion level and University of California Los Angeles score (P = .016). Conclusion: We conclude that serum metal level elevation may occur secondary to metal debris resulting from corrosion of the index MDM THA. This potential risk should be included in the decision-making process when dealing with revision arthroplasty in young and active patients.

Patients Following Revision Total Hip Arthroplasty With Modular Dual Mobility Components and Cobalt-Chromium Inner Metal Head are at Risk of Increased Serum Metal Ion Levels / Civinini R.; Cozzi Lepri A.; Carulli C.; Matassi F.; Villano M.; Innocenti M.. - In: THE JOURNAL OF ARTHROPLASTY. - ISSN 0883-5403. - STAMPA. - 36:(2020), pp. 1-5. [10.1016/j.arth.2020.02.050]

Patients Following Revision Total Hip Arthroplasty With Modular Dual Mobility Components and Cobalt-Chromium Inner Metal Head are at Risk of Increased Serum Metal Ion Levels

Civinini R.
;
Cozzi Lepri A.;Carulli C.;Matassi F.;Innocenti M.
2020

Abstract

Background: Modular dual-mobility (MDM) total hip arthroplasty (THA) is designed with a cobalt-chromium liner inserted into a titanium acetabular component. The purpose of this study is to investigate the potential risks for fretting corrosion at this junction, by measuring serum metal ions, after MDM acetabular revision. Methods: Thirty-seven patients with well-functioning revision THAs participated in a cross-sectional study at mean 5.1 (2-10) years after surgery. All received a trabecular titanium MDM acetabular component. The serum levels of cobalt and chromium were measured using mass spectrometry. Results: The mean values of chromium and cobalt were 2.08 μg/L (95% confidence interval 0.9-3.2, range 0.02-11.8) and 1.99 μg/L (95% confidence interval 0.81-3.17, range 0.07-16.05), respectively.Eleven patients (29, 7%) had ion levels above the normal range, with 6 (16.2%) above 5 μg/L and 5 (13.5%) between 1 and 5 μg/L. A significant correlation was found between an elevated serum metal ion level and University of California Los Angeles score (P = .016). Conclusion: We conclude that serum metal level elevation may occur secondary to metal debris resulting from corrosion of the index MDM THA. This potential risk should be included in the decision-making process when dealing with revision arthroplasty in young and active patients.
2020
36
1
5
Civinini R.; Cozzi Lepri A.; Carulli C.; Matassi F.; Villano M.; Innocenti M.
File in questo prodotto:
File Dimensione Formato  
metal ions rTHA.pdf

Accesso chiuso

Tipologia: Pdf editoriale (Version of record)
Licenza: Tutti i diritti riservati
Dimensione 318.9 kB
Formato Adobe PDF
318.9 kB Adobe PDF   Richiedi una copia

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/1189224
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 26
  • ???jsp.display-item.citation.isi??? 25
social impact